Preliminary Operations; Getting Started - Thermo Scientific TRACE 1600 Hardware Manual

Gas chromatographs
Table of Contents

Advertisement

Preliminary Operations

Getting Started

Thermo Scientific
Figure 580. External Modules Housing
Before starting, the following preliminary operation must be carried out.
1. Put the GC in standby condition.
2. Cool the oven, injectors, and detectors to room temperature.
Note By pressing the Maintenance button, the GC cool down is automatically
carried out.
3. Close the gas supplies.
4. Power off the GC.
a. Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector on the back of the GC, and
from the wall outlet.
5. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
To install and connect the Auxiliary Gas Interface and the Auxiliary Gas Module, see the
following sections:
"Installing the Auxiliary Gas Interface on the Right Wall of the Oven"
"Installing the Auxiliary Gas Interface on the Oven for HRMS"
"Installing and Connecting the Auxiliary Gas Module"
Adding an Auxiliary Gas System
External Modules Housing
on
on
page 568
TRACE 1600/1610 Hardware Manual
10
Adding Systems
on
page 562
page 565
561