:SAVE Commands

:SAVE:WAVeform:SEGMented

Command Syntax
:SAVE:WAVeform:SEGMented <option>
<option> ::= {ALL | CURRent}
When segmented memory is used for acquisitions, the
:SAVE:WAVeform:SEGMented command specifies which segments are
included when the waveform is saved:
• ALL — all acquired segments are saved.
• CURRent — only the currently selected segment is saved.
Query Syntax
:SAVE:WAVeform:SEGMented?
The :SAVE:WAVeform:SEGMented? query returns the current segmented
waveform save option setting.
Return Format
<option><NL>
<option> ::= {ALL | CURR}