:MEASure:WINDow

Command Syntax
:MEASure:WINDow <type>
<type> ::= {MAIN | ZOOM | AUTO | GATE}
The :MEASure:WINDow command lets you choose whether measurements
are made in the Main window portion of the display, the Zoom window
portion of the display (when the zoomed time base is displayed), or gated
by the X1 and X2 cursors.
• MAIN — the measurement window is the Main window.
• ZOOM — the measurement window is the lower, Zoom window.
• AUTO — when the zoomed time base is displayed, the measurement is
• GATE — the measurement window is between the X1 and X2 cursors.
Query Syntax
:MEASure:WINDow?
The :MEASure:WINDow? query returns the current measurement window
setting.
Return Format
<type><NL>
<type> ::= {MAIN | ZOOM | AUTO | GATE}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
attempted in the lower, Zoom window; if it cannot be made there, or if
the zoomed time base is not displayed, the Main window is used.
When the zoomed time base is displayed, the X1 and X2 cursors in the
Zoom window portion of the display are used.