Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual page 293

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

When the education kit is licensed (Option EDU), you can output
demonstration signals on the oscilloscope's Demo 1 and Demo 2 terminals.
See
Table 87 :DEMO Commands Summary
Command
:DEMO:FUNCtion
<signal> (see
page
294)
:DEMO:FUNCtion:PHASe:
PHASe <angle> (see
page
298)
:DEMO:OUTPut {{0 |
OFF} | {1 | ON}} (see
page
299)
Introduction to
The :DEMO subsystem provides commands to output demonstration signals
:DEMO
on the oscillosope's Demo 1 and Demo 2 terminals.
Commands
Reporting the Setup
Use :DEMO? to query setup information for the DEMO subsystem.
Return Format
The following is a sample response from the :DEMO? query. In this case,
the query was issued following the *RST command.
:DEMO:FUNC SIN;OUTP 0
Agilent InfiniiVision 4000 X-Series Oscilloscopes
Programmer's Guide
12
:DEMO Commands
"Introduction to :DEMO Commands"
Query
:DEMO:FUNCtion? (see
page
297)
:DEMO:FUNCtion:PHASe:
PHASe? (see
page
:DEMO:OUTPut? (see
page
299)
on page 293.
Options and Query Returns
<signal> ::= {SINusoid | NOISy |
PHASe | RINGing | SINGle | AM |
CLK | GLITch | BURSt | MSO | RUNT
| TRANsition | RFBurst | SHOLd |
LFSine | FMBurst | ETE | CAN |
LIN | UART | I2C | SPI | I2S |
CANLin | ARINc | FLEXray | MIL |
MIL2 | USB | NMONotonic}
<angle> ::= angle in degrees from
298)
0 to 360 in NR3 format
{0 | 1}