:MTESt:SCALe:Y1

Command Syntax
:MTESt:SCALe:Y1 <y1_value>
<y1_value> ::= Y1 value in NR3 format
The :MTESt:SCALe:Y1 command defines where Y=0 in the coordinate
system for mask testing. All Y values of vertices in the coordinate system
are defined with respect to the boundaries set by SCALe:Y1 and SCALe:Y2
according to the equation:
Y = (Y * (Y2 - Y1)) + Y1
Thus, if you set Y1 to 100 mV, and Y2 to 1 V, a Y value of 0.100 in a
vertex is at 190 mV.
The Y1 value is a voltage value specifying the point at which Y=0.
Query Syntax
:MTESt:SCALe:Y1?
The :MTESt:SCALe:Y1? query returns the current setting of the Y1 marker.
Return Format
<y1_value><NL>
<y1_value> ::= Y1 value in NR3 format
See Also