22
:MEASure Power Commands

:MEASure:ONTime

Command Syntax
:MEASure:ONTime [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:ONTime command installs an "on time" measurement on
screen.
Turn on time measures the difference of time between when the input AC
Voltage first rises to 10% of its maximum amplitude to the time when the
output DC Voltage rises to 90% of its maximum amplitude.
The <source1> parameter is the AC Voltage and the <source2> parameter is
the DC Voltage. These sources can also be specified by the
:MEASure:SOURce command.
Query Syntax
:MEASure:ONTime? [<source1>][,<source2>]
The :MEASure:ONTime query returns the measured turn off time.
Return Format
<return_value><NL>
<return_value> ::= the time in seconds in NR3 format
See Also
498
(see
page
1208)
":MEASure:SOURce"
":POWer:ONOFf:TEST"
":POWer:ONOFf:APPLy"