:SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command sets the
number of 8- bit bytes in the CAN data string. The number of bytes in the
string can be anywhere from 1 bytes to 8 bytes (64 bits). The value for
these bytes is set by the :SBUS<n>:CAN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth? query returns the
current CAN data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors
See Also
682
(see
page
1208)
"- 241, Hardware missing"
"Introduction to :TRIGger Commands"
":SBUS<n>:CAN:TRIGger:PATTern:DATA"
":SBUS<n>:CAN:SOURce"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 1169
on page 935
on page 681