:WGEN<w> Commands
:WGEN<w>:OUTPut:LOAD
Command Syntax
:WGEN<w>:OUTPut:LOAD <impedance>
<w> ::= 1 or 2 in NR1 format
<impedance> ::= {ONEMeg | FIFTy}
The :WGEN<w>:OUTPut:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However,
the output load selection lets the waveform generator display the correct
amplitude and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the
displayed amplitude and offset levels will be incorrect.
Query Syntax
:WGEN<w>:OUTPut:LOAD?
The :WGEN<w>:OUTPut:LOAD? query returns the current expected output
load impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}