:TRIGger Commands
:TRIGger:ZONE<n>:STATe
Command Syntax
:TRIGger:ZONE<n>:STATe <on_off>
<n> ::= {1 | 2}
<on_off> ::= {{0 | OFF} | {1 | ON}}
<n> ::= 1-2 in NR1 format
The :TRIGger:ZONE<n>:STATe command sets the state for Zone 1 or Zone
2.
• When a zone's state is on, and the Zone Trigger feature is on (see
• When the Zone Trigger feature is off, no zones are being used to qualify
Note that :TRIGger:ZONE:STATe mimics the behavior of the [Zone] key on
the front panel, and :TRIGger:ZONE<n>:STATe mimics the behavior of the
Zone 1 On and Zone 2 On softkeys. At least one zone's state must be on for
the Zone Trigger feature (:TRIGger:ZONE:STATe) to be on. When the states
of both individual zones are turned off, Zone Trigger is automatically
turned off. In this case, when Zone Trigger is turned back on Zone 1 is
forced to on. Otherwise, if at least one zone was on when Zone Trigger
was turned off, the same configuration of individual zone on/off states will
be restored when Zone Trigger is turned back on.
Query Syntax
:TRIGger:ZONE<n>:STATe?
The :TRIGger:ZONE<n>:STATe? query returns the state of Zone 1 or Zone
2.
Return Format
<on_off><NL>
<on_off> ::= {0 | 1}
See Also
1018
(see
page
1208)
":TRIGger:ZONE:STATe"
to qualify the trigger if it is not invalid or off- screen (see
":TRIGger:ZONE<n>:VALidity"
the trigger, regardless of their individual states.
":TRIGger:ZONE:STATe"
":TRIGger:ZONE<n>:VALidity"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 1014), that zone is actively being used
on page 1017).
on page 1014
on page 1017

Advertisement

Table of Contents
loading

Table of Contents