28
:SBUS<n> Commands
:SBUS<n>:UART:BAUDrate
Command Syntax
:SBUS<n>:UART:BAUDrate <baudrate>
<baudrate> ::= integer from 100 to 8000000
The :SBUS<n>:UART:BAUDrate command selects the bit rate (in bps) for
the serial decoder and/or trigger when in UART mode. The baud rate can
be set from 100 b/s to 8 Mb/s.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:SBUS<n>:UART:BAUDrate?
The :SBUS<n>:UART:BAUDrate? query returns the current UART baud rate
setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 100 to 8000000
See Also
774
(see
page
1208)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 944
":SBUS<n>:UART:TRIGger:TYPE"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 935
on page 790