28
:SBUS<n> Commands
:SBUS<n>:SPI:SOURce:CLOCk
Command Syntax
:SBUS<n>:SPI:SOURce:CLOCk <source>
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d>} for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :SBUS<n>:SPI:SOURce:CLOCk command sets the source for the SPI
serial clock.
Query Syntax
:SBUS<n>:SPI:SOURce:CLOCk?
The :SBUS<n>:SPI:SOURce:CLOCk? query returns the current source for
the SPI serial clock.
Return Format
<source><NL>
See Also
760
(see
page
1208)
"Introduction to :TRIGger Commands"
":SBUS<n>:SPI:CLOCk:SLOPe"
":SBUS<n>:SPI:CLOCk:TIMeout"
":SBUS<n>:SPI:SOURce:FRAMe"
":SBUS<n>:SPI:SOURce:MOSI"
":SBUS<n>:SPI:SOURce:MISO"