:MTESt:AMASk:UNITs
Command Syntax
:MTESt:AMASk:UNITs <units>
<units> ::= {CURRent | DIVisions}
The :MTESt:AMASk:UNITs command alters the way the mask test
subsystem interprets the tolerance parameters for automasking as defined
by :MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta commands.
• CURRent — the mask test subsystem uses the units as set by the
• DIVisions — the mask test subsystem uses the graticule as the
Query Syntax
:MTESt:AMASk:UNITs?
The :MTESt:AMASk:UNITs query returns the current measurement units
setting for the mask test automask feature.
Return Format
<units><NL>
<units> ::= {CURR | DIV}
See Also
•
•
•
•
•
•
Example Code
•
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
:CHANnel<n>:UNITs command, usually time for X and voltage for Y.
measurement system, so tolerance settings are specified as parts of a
screen division. The mask test subsystem maintains separate XDELta
and YDELta settings for CURRent and DIVisions. Thus, XDELta and
YDELta are not converted to new values when the UNITs setting is
changed.
"Introduction to :MTESt Commands"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:YDELta"
":CHANnel<n>:UNITs"
":MTESt:AMASk:SOURce"
":MTESt:SOURce"
on page 539
"Example Code"
on page 509
on page 509
on page 516
on page 517
on page 281
on page 514