Demo:function - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

12
:DEMO Commands

:DEMO:FUNCtion

Command Syntax
:DEMO:FUNCtion <signal>
<signal> ::= {SINusoid | NOISy | PHASe | RINGing | SINGle | AM | CLK
The :DEMO:FUNCtion command selects the type of demo signal:
Demo Signal
Function
SINusoid
NOISy
PHAse
RINGing
SINGle
AM
CLK
GLITch
BURSt
294
(see
page
1208)
| GLITch | BURSt | MSO | RUNT | TRANsition | RFBurst
| SHOLd | LFSine | FMBurst | ETE | CAN | LIN | UART
| I2C | SPI | I2S | CANLin | ARINc | FLEXray | MIL
| MIL2 | USB | NMONotonic}
Demo 1 Terminal
5 MHz sine wave @ ~ 6 Vpp, 0 V
offset
1 kHz sine wave @ ~ 2.4 Vpp, 0.0 V
offset, with ~ 0.5 Vpp of random
noise added
1 kHz sine wave @ 2.4 Vpp, 0.0 V
offset
500 kHz digital pulse @ ~ 3 Vpp, 1.5 V
offset, and ~500 ns pulse width with
ringing
~500 ns wide digital pulse with
ringing @ ~ 3 Vpp, 1.5 V offset
Press the front panel Set Off
Single-Shot softkey to cause the
selected single-shot signal to be
output.
26 kHz sine wave, ~ 7 Vpp, 0 V offset
3.6 MHz clock @ ~2 Vpp, 1 V offset,
with infrequent glitch (1 glitch per
1,000,000 clocks)
Burst of 6 digital pulses (plus
infrequent glitch) that occurs once
every 80 µs @ ~3.6 Vpp, ~1.8 V offset
Burst of digital pulses that occur
every 50 µs @ ~ 3.6 Vpp, ~1.5 V offset
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
Demo 2 Terminal
Off
Off
1 kHz sine wave @ 2.4 Vpp, 0.0 V
offset , phase shifted by the amount
entered using the
":DEMO:FUNCtion:PHASe:PHASe"
on page 298 command
Off
Off
Amplitude modulated signal, ~ 3 Vpp,
0 V offset, with ~13 MHz carrier and
sine envelope
Off
Off
Off

Advertisement

Table of Contents
loading