Sbus<N>:Lin:syncbreak - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

28
:SBUS<n> Commands
:SBUS<n>:LIN:SYNCbreak
Command Syntax
:SBUS<n>:LIN:SYNCbreak <value>
<value> ::= integer = {11 | 12 | 13}
The :SBUS<n>:LIN:SYNCbreak command sets the length of the LIN sync
break to be greater than or equal to 11, 12, or 13 clock lengths. The sync
break is the idle period in the bus activity at the beginning of each packet
that distinguishes one information packet from the previous one.
Query Syntax
:SBUS<n>:LIN:SYNCbreak?
The :SBUS<n>:LIN:SYNCbreak? query returns the current LIN sync break
setting.
Return Format
<value><NL>
<value> ::= {11 | 12 | 13}
See Also
740
(see
page
1208)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 944
":SBUS<n>:LIN:SIGNal:DEFinition"
":SBUS<n>:LIN:SOURce"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 935
on page 1161
on page 738