Measure:ripple - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

22
:MEASure Power Commands

:MEASure:RIPPle

Command Syntax
:MEASure:RIPPle [<source>]
<source> ::= {CHANnel<n>| FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:RIPPle command installs an output ripple measurement on
screen.
The <source> parameter is the channel probing the output voltage. This
source can also be specified by the :MEASure:SOURce command.
Output ripple is: Vmax - Vmin.
Query Syntax
:MEASure:RIPPle? [<source>]
The :MEASure:RIPPle query returns the measured output ripple.
Return Format
<return_value><NL>
<return_value> ::= the output ripple value in NR3 format
See Also
504
(see
page
1208)
":MEASure:SOURce"
":POWer:RIPPle:APPLy"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 458
on page 580

Advertisement

Table of Contents
loading

Table of Contents