28
:SBUS<n> Commands
:SBUS<n>:USB:TRIGger:PID:SPECial
Command Syntax
:SBUS<n>:USB:TRIGger:PID:SPECial <pid>
<pid> ::= {PING | PRE | ERR | SPLit}
The :SBUS<n>:USB:TRIGger:PID:SPECial command specifies the USB
special PID to trigger on for the selected bus.
The specified PID does not include the PID check value, which is specified
using the :SBUS<n>:USB:TRIGger:PIDCheck command.
The SPECial USB trigger mode can be selected using the
:SBUS<n>:USB:TRIGger command.
Query Syntax
:SBUS<n>:USB:TRIGger:PID:SPECial?
The :SBUS<n>:USB:TRIGger:PID:SPECial? query returns the specified
special PID.
Return Format
<pid><NL>
<pid> ::= {PING | PRE | ERR | SPL}
See Also
812
(see
page
1208)
":SBUS<n>:USB:TRIGger:PCHeck"
":SBUS<n>:USB:TRIGger"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 809
on page 800