:WGEN<w>:MODulation:AM:DEPTh
Command Syntax
:WGEN<w>:MODulation:AM:DEPTh <percent>
<w> ::= 1 in NR1 format
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
The :WGEN<w>:MODulation:AM:DEPTh command specifies the amount of
amplitude modulation.
AM Depth refers to the portion of the amplitude range that will be used
by the modulation. For example, a depth setting of 80% causes the output
amplitude to vary from 10% to 90% (90% – 10% = 80%) of the original
amplitude as the modulating signal goes from its minimum to maximum
amplitude.
Query Syntax
:WGEN<w>:MODulation:AM:DEPTh?
The :WGEN<w>:MODulation:AM:DEPTh? query returns the AM depth
percentage setting.
Return Format
<percent><NL>
<percent> ::= AM depth percentage from 0% to 100% in NR1 format
See Also
•
•
•
•
•
•
•
•
•
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
":WGEN<w>:MODulation:AM:FREQuency"
":WGEN<w>:MODulation:FM:DEViation"
":WGEN<w>:MODulation:FM:FREQuency"
":WGEN<w>:MODulation:FSKey:FREQuency"
":WGEN<w>:MODulation:FSKey:RATE"
":WGEN<w>:MODulation:FUNCtion"
":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry"
":WGEN<w>:MODulation:STATe"
":WGEN<w>:MODulation:TYPE"
:WGEN<w> Commands
on page 1074
on page 1075
on page 1076
on page 1077
on page 1078
on page 1079
on page 1082
on page 1083
34
on page 1080
1073