33
:WAVeform Commands
Table 144 :WAVeform Commands Summary (continued)
Command
:WAVeform:POINts:MODE
<points_mode> (see
page
1034)
n/a
n/a
n/a
:WAVeform:SOURce
<source> (see
page
1041)
:WAVeform:SOURce:SUBS
ource <subsource>
(see
page
1045)
n/a
1020
Query
:WAVeform:POINts:MODE
? (see
page
1034)
:WAVeform:PREamble?
(see
page
1036)
:WAVeform:SEGMented:C
OUNt? (see
page
1039)
:WAVeform:SEGMented:T
TAG? (see
page
1040)
:WAVeform:SOURce?
(see
page
1041)
:WAVeform:SOURce:SUBS
ource? (see
page
1045)
:WAVeform:TYPE? (see
page
1046)
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
Options and Query Returns
<points_mode> ::= {NORMal |
MAXimum | RAW}
<preamble_block> ::= <format
NR1>, <type NR1>,<points
NR1>,<count NR1>, <xincrement
NR3>, <xorigin NR3>, <xreference
NR1>,<yincrement NR3>, <yorigin
NR3>, <yreference NR1>
<format> ::= an integer in NR1
format:
• 0 for BYTE format
• 1 for WORD format
• 2 for ASCii format
<type> ::= an integer in NR1
format:
• 0 for NORMal type
• 1 for PEAK detect type
• 3 for AVERage type
• 4 for HRESolution type
<count> ::= Average count, or 1
if PEAK detect type or NORMal; an
integer in NR1 format
<count> ::= an integer from 2 to
1000 in NR1 format (with Option
SGM)
<time_tag> ::= in NR3 format
(with Option SGM)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> | SBUS} for
DSO models
<source> ::= {CHANnel<n> | POD{1
| 2} | BUS{1 | 2} | FUNCtion<m> |
MATH<m> | SBUS} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1-4 in NR1 format
<subsource> ::= {{SUB0 | RX |
MOSI} | {SUB1 | TX | MISO}}
<return_mode> ::= {NORM | PEAK |
AVER | HRES}