General :TRIGger Commands

Table 131 General :TRIGger Commands Summary
Command
:TRIGger:FORCe (see
page
938)
:TRIGger:HFReject {{0
| OFF} | {1 | ON}}
(see
page
939)
:TRIGger:HOLDoff
<holdoff_time> (see
page
940)
:TRIGger:LEVel:ASETup
(see
page
941)
:TRIGger:LEVel:HIGH
<level>, <source>
(see
page
942)
:TRIGger:LEVel:LOW
<level>, <source>
(see
page
943)
:TRIGger:MODE <mode>
(see
page
944)
:TRIGger:NREJect {{0
| OFF} | {1 | ON}}
(see
page
945)
:TRIGger:SWEep
<sweep> (see
page
946)
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
Query
n/a
:TRIGger:HFReject?
(see
page
939)
:TRIGger:HOLDoff?
(see
page
940)
n/a
:TRIGger:LEVel:HIGH?
<source> (see
page
942)
:TRIGger:LEVel:LOW?
<source> (see
page
943)
:TRIGger:MODE? (see
page
944)
:TRIGger:NREJect?
(see
page
945)
:TRIGger:SWEep? (see
page
946)
:TRIGger Commands
Options and Query Returns
n/a
{0 | 1}
<holdoff_time> ::= 60 ns to 10 s
in NR3 format
n/a
<level> ::= .75 x full-scale
voltage from center screen in NR3
format.
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
<level> ::= .75 x full-scale
voltage from center screen in NR3
format.
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
<mode> ::= {EDGE | GLITch |
PATTern | TV | DELay | EBURst |
OR | RUNT | SHOLd | TRANsition |
SBUS{1 | 2}}
<return_value> ::= {<mode> |
<none>}
<none> ::= query returns "NONE"
if the :TIMebase:MODE is ROLL or
XY
{0 | 1}
<sweep> ::= {AUTO | NORMal}