:MEASure:PERiod

Command Syntax
:MEASure:PERiod [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion<m> | MATH<m>
<digital channels> ::= DIGital<d> for the MSO models
<d> ::= 0 to (# digital channels - 1) in NR1 format
<n> ::= 1 to (# of analog channels) in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:PERiod command installs a screen measurement and starts
the period measurement. If the optional source parameter is specified, the
current source is modified.
This command is not available if the source is FFT (Fast Fourier Transform).
N O T E
Query Syntax
:MEASure:PERiod? [<source>]
The :MEASure:PERiod? query measures and outputs the period of the
cycle closest to the trigger reference on the screen. The period is measured
at the midpoint of the upper and lower thresholds.
IF the edge closest to the trigger reference on screen is rising:
THEN period = (time at trailing rising edge - time at leading rising edge)
ELSE period = (time at trailing falling edge - time at leading falling edge)
Return Format
<value><NL>
<value> ::= waveform period in seconds in NR3 format
See Also
Example Code
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
| WMEMory<r>}
"Introduction to :MEASure Commands"
":MEASure:SOURce"
":MEASure:NWIDth"
":MEASure:PWIDth"
on page 451
":MEASure:FREQuency"
"Example Code"
on page 459
on page 420
on page 458
on page 443
on page 440
21
:MEASure Commands
447

Advertisement

Table of Contents
loading

Table of Contents