:MARKer Commands
:MARKer:MODE
Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual | WAVeform | BINary | HEX}
The :MARKer:MODE command sets the cursors mode:
• OFF — removes the cursor information from the display.
• MANual — enables manual placement of the X and Y cursors.
• MEASurement — cursors track the most recent measurement.
• WAVeform — the Y1 cursor tracks the voltage value at the X1 cursor of
• BINary — logic levels of displayed waveforms at the current X1 and X2
• HEX — logic levels of displayed waveforms at the current X1 and X2
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN | WAV | BIN | HEX}
See Also
394
(see
page
1208)
If the front- panel cursors are off, or are set to the front- panel Hex or
Binary mode, setting :MARKer:MODE MANual will put the cursors in
the front- panel Normal mode.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement
source (:MEASure:SOURce). Setting the measurement source remotely
always sets the marker sources.
the waveform specified by the X1Y1source, and the Y2 cursor does the
same for the X2 cursor and its X2Y2source.
cursor positions are displayed in the Cursor sidebar dialog in binary.
cursor positions are displayed in the Cursor sidebar dialog in
hexadecimal.
"Introduction to :MARKer Commands"
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MEASure:SOURce"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:Y1Position"
":MARKer:Y2Position"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 392
on page 396
on page 398
on page 458
on page 395
on page 397
on page 402
on page 403

Advertisement

Table of Contents
loading

Table of Contents