:FUNCtion<m>:OFFSet
Command Syntax
:FUNCtion<m>:OFFSet <offset>
<m> ::= 1-4 in NR1 format
<offset> ::= the value at center screen in NR3 format.
The :FUNCtion<m>:OFFSet command sets the voltage or vertical value
represented at center screen for the selected function. The range of legal
values is generally +/- 10 times the current scale of the selected function,
but will vary by function. If you set the offset to a value outside of the
legal range, the offset value is automatically set to the nearest legal value.
The :FUNCtion<m>:OFFset command is equivalent to the :FUNCtion<m>:REFerence
N O T E
command.
Query Syntax
:FUNCtion<m>:OFFSet?
The :FUNCtion<m>:OFFSet? query outputs the current offset value for the
selected function.
Return Format
<offset><NL>
<offset> ::= the value at center screen in NR3 format.
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :FUNCtion<m> Commands"
":FUNCtion<m>:RANGe"
":FUNCtion<m>:REFerence"
":FUNCtion<m>:SCALe"
on page 361
on page 362
on page 363
:FUNCtion<m> Commands
on page 340