29
:SEARch Commands

:SEARch:SERial:USB:PID:TOKen

(see
page
1208)
Command Syntax
:SEARch:SERial:USB:PID:TOKen <pid>
<pid> ::= {OUT | IN | SETup | SOF}
The :SEARch:SERial:USB:PID:TOKen command specifies the USB token PID
to find.
The TOKen USB search mode can be selected using the
:SEARch:SERial:USB:MODE command.
Query Syntax
:SEARch:SERial:USB:PID:TOKen?
The :SEARch:SERial:USB:PID:TOKen? query returns the specified token
PID.
Return Format
<pid><NL>
<pid> ::= {OUT | IN | SETup | SOF}
See Also
":SEARch:SERial:USB:MODE"
on page 893