:EXTernal:PROBe

Command Syntax
:EXTernal:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
The :EXTernal:PROBe command specifies the probe attenuation factor for
the external trigger. The probe attenuation factor may be 0.1 to 1000. This
command does not change the actual input sensitivity of the oscilloscope.
It changes the reference constants for scaling the display factors and for
setting trigger levels.
If an AutoProbe probe is connected to the oscilloscope, the attenuation
value cannot be changed from the sensed value. Attempting to set the
oscilloscope to an attenuation value other than the sensed value produces
an error.
Query Syntax
:EXTernal:PROBe?
The :EXTernal:PROBe? query returns the current probe attenuation factor
for the external trigger.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :EXTernal Trigger Commands"
":EXTernal:RANGe"
on page 334
"Introduction to :TRIGger Commands"
":CHANnel<n>:PROBe"
on page 935
on page 273
:EXTernal Trigger Commands
on page 331