Waveform:unsigned - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:WAVeform:UNSigned

Command Syntax
:WAVeform:UNSigned <unsigned>
<unsigned> ::= {{0 | OFF} | {1 | ON}}
The :WAVeform:UNSigned command turns unsigned mode on or off for the
currently selected waveform. Use the WAVeform:UNSigned command to
control whether data values are sent as unsigned or signed integers. This
command can be used to match the instrument's internal data type to the
data type used by the programming language. This command has no effect
if the data format is ASCii.
If :WAVeform:SOURce is set to POD1, POD2, BUS1, or BUS2,
WAVeform:UNSigned must be set to ON.
Query Syntax
:WAVeform:UNSigned?
The :WAVeform:UNSigned? query returns the status of unsigned mode for
the currently selected waveform.
Return Format
<unsigned><NL>
<unsigned> ::= {0 | 1}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :WAVeform Commands"
":WAVeform:SOURce"
on page 1021
on page 1041
33
:WAVeform Commands
1047