32
:TRIGger Commands
:TRIGger:PATTern:FORMat
Command Syntax
:TRIGger:PATTern:FORMat <base>
<base> ::= {ASCii | HEX}
The :TRIGger:PATTern:FORMat command sets the entry (and query)
number base used by the :TRIGger:PATTern command. The default <base>
is ASCii.
Query Syntax
:TRIGger:PATTern:FORMat?
The :TRIGger:PATTern:FORMat? query returns the currently set number
base for pattern trigger patterns.
Return Format
<base><NL>
<base> ::= {ASC | HEX}
See Also
•
•
980
(see
page
1208)
"Introduction to :TRIGger Commands"
":TRIGger:PATTern"
on page 978
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 935