13
:DIGital<d> Commands
:DIGital<d>:LABel
Command Syntax
:DIGital<d>:LABel <string>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<string> ::= any series of 10 or less characters as quoted ASCII string.
The :DIGital<d>:LABel command sets the channel label to the string that
follows. Setting a label for a channel also adds the name to the label list
in non- volatile memory (replacing the oldest label in the list).
This command is only valid for the MSO models.
N O T E
Label strings are 10 characters or less, and may contain any commonly used ASCII
N O T E
characters. Labels with more than 10 characters are truncated to 10 characters.
Query Syntax
:DIGital<d>:LABel?
The :DIGital<d>:LABel? query returns the name of the specified channel.
Return Format
<label string><NL>
<label string> ::= any series of 10 or less characters as a quoted
See Also
304
(see
page
1208)
ASCII string.
"Introduction to :DIGital<d> Commands"
":CHANnel<n>:LABel"
":DISPlay:LABList"
on page 318
":BUS<n>:LABel"
on page 250
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide