:SBUS<n>:USB:SPEed
Command Syntax
:SBUS<n>:USB:SPEed <speed>
<speed> ::= {LOW | FULL | HIGH}
The :SBUS<n>:USB:SPEed command specifies the speed of the USB
interface for the selected bus.
Query Syntax
:SBUS<n>:USB:SPEed?
The :SBUS<n>:USB:SPEed? query returns the speed setting.
Return Format
<speed><NL>
<speed> ::= {LOW | FULL | HIGH}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
":SBUS<n>:USB:SOURce:DPLus"
":SBUS<n>:USB:SOURce:DMINus"
":SBUS<n>:USB:SOURce:DIFFerential"
:SBUS<n> Commands
on page 797