Power:signals:vsteady:onoff:off - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:POWer:SIGNals:VSTeady:ONOFf:OFF

Command Syntax
:POWer:SIGNals:VSTeady:ONOFf:OFF <value>[suffix]
<value> ::= Expected steady state output Voltage value in NR3 format
[suffix] ::= {V | mV}
The :POWer:SIGNals:VSTeady:ONOFf:OFF command specifies the expected
steady state output DC voltage of the power supply for turn off analysis.
Query Syntax
:POWer:SIGNals:VSTeady:ONOFf:OFF?
The :POWer:SIGNals:VSTeady:ONOFf:OFF query returns the expected steady
state voltage setting.
Return Format
<value><NL>
<value> ::= Expected steady state output Voltage value in NR3 format
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
":POWer:ONOFf:APPLy"
":POWer:SIGNals:AUTosetup"
":POWer:SIGNals:DURation:ONOFf:OFF"
":POWer:SIGNals:IEXPected"
":POWer:SIGNals:OVERshoot"
":POWer:SIGNals:VMAXimum:ONOFf:OFF"
":POWer:SIGNals:SOURce:CURRent<i>"
":POWer:SIGNals:SOURce:VOLTage<i>"
on page 571
on page 581
on page 586
on page 590
on page 591
on page 593
on page 598
on page 599
25
:POWer Commands
595

Advertisement

Table of Contents
loading

Table of Contents