:SBUS<n>:UART:TRIGger:DATA
Command Syntax
:SBUS<n>:UART:TRIGger:DATA <value>
<value> ::= 8-bit integer from 0-255 (0x00-0xff) in decimal,
<hexadecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<binary> ::= #Bnn...n where n ::= {0 | 1} for binary
<quoted_string> ::= any of the 128 valid 7-bit ASCII characters
The :SBUS<n>:UART:TRIGger:DATA command selects the data byte value
(0x00 to 0xFF) for the trigger QUALifier when in UART mode. The data
value is used when one of the RD or TD trigger types is selected.
When entering an ASCII character via the quoted string, it must be one of
the 128 valid characters (case- sensitive): "NUL", "SOH", "STX", "ETX",
"EOT", "ENQ", "ACK", "BEL", "BS", "HT", "LF", "VT", "FF", "CR", "SO","SI",
"DLE", "DC1", "DC2", "DC3", "DC4", "NAK", "SYN", "ETB", "CAN", "EM", "SUB",
"ESC", "FS","GS", "RS", "US", "SP", "!", "\"", "#", "$", "%","&", "\'", "(", ")", "*", "+",
",", "- ", ".", "/","0", "1", "2", "3", "4", "5", "6", "7", "8", "9",":", ";", "<", "=", ">", "?",
"@", "A", "B", "C","D", "E", "F", "G", "H", "I", "J", "K", "L", "M","N", "O", "P", "Q",
"R", "S", "T", "U", "V", "W", "X", "Y", "Z","[", "\\", "]", "^", "_", "`", "a", "b", "c",
"d", "e", "f", "g", "h", "i", "j", "k", "l", "m", "n", "o", "p", "q", "r", "s", "t", "u", "v",
"w", "x", "y", "z", "{", "|", "}", "~", or "DEL".
Query Syntax
:SBUS<n>:UART:TRIGger:DATA?
The :SBUS<n>:UART:TRIGger:DATA? query returns the current UART
trigger data value.
Return Format
<value><NL>
<value> ::= 8-bit integer in decimal from 0-255
See Also
•
•
•
•
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
<hexadecimal>, <binary>, or <quoted_string> format
(or standard abbreviations)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 944
":SBUS<n>:UART:TRIGger:BASE"
":SBUS<n>:UART:TRIGger:TYPE"
:SBUS<n> Commands
on page 935
on page 785
on page 790
28
787