Sbus<N>:Uart:trigger:base - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:SBUS<n>:UART:TRIGger:BASE
Command Syntax
:SBUS<n>:UART:TRIGger:BASE <base>
<base> ::= {ASCii | HEX}
The :SBUS<n>:UART:TRIGger:BASE command sets the front panel
UART/RS232 trigger setup data selection option:
• ASCii — front panel data selection is from ASCII values.
• HEX — front panel data selection is from hexadecimal values.
The :SBUS<n>:UART:TRIGger:BASE setting does not affect the
:SBUS<n>:UART:TRIGger:DATA command which can always set data values
using ASCII or hexadecimal values.
The :SBUS<n>:UART:TRIGger:BASE command is independent of the
N O T E
:SBUS<n>:UART:BASE command which affects decode and Lister only.
Query Syntax
:SBUS<n>:UART:TRIGger:BASE?
The :SBUS<n>:UART:TRIGger:BASE? query returns the current UART base
setting.
Return Format
<base><NL>
<base> ::= {ASC | HEX}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 944
":SBUS<n>:UART:TRIGger:DATA"
:SBUS<n> Commands
on page 935
on page 787
28