:LISTer:DISPlay
Command Syntax
:LISTer:DISPlay <value>
<value> ::= {{OFF | 0} | {SBUS1 | ON | 1} | {SBUS2 | 2} | ALL}
The :LISTer:DISPlay command configures which of the serial buses to
display in the Lister, or whether the Lister is off. "ON" or "1" is the same
as "SBUS1".
When set to "ALL", the decode information for different buses is
interleaved in time.
Serial bus decode must be on before it can be displayed in the Lister.
Query Syntax
:LISTer:DISPlay?
The :LISTer:DISPlay? query returns the Lister display setting.
Return Format
<value><NL>
<value> ::= {OFF | SBUS1 | SBUS2 | ALL}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :LISTer Commands"
":SBUS<n>:DISPlay"
on page 648
":LISTer:DATA"
on page 388
:LISTer Commands
on page 387