25
:POWer Commands

:POWer:SIGNals:CYCLes:HARMonics

Command Syntax
:POWer:SIGNals:CYCLes:HARMonics <count>
<count> ::= integer in NR1 format
Legal values are 1 to 100.
The :POWer:SIGNals:CYCLes:HARMonics command specifies the number of
cycles to include in the current harmonics analysis.
Query Syntax
:POWer:SIGNals:CYCLes:HARMonics?
The :POWer:SIGNals:CYCLes:HARMonics query returns the number of
cycles currently set.
Return Format
<count><NL>
<count> ::= integer in NR1 format
See Also
582
(see
page
1208)
":POWer:HARMonics:DISPlay"
":POWer:HARMonics:APPLy"
":POWer:SIGNals:AUTosetup"
":POWer:SIGNals:IEXPected"
":POWer:SIGNals:OVERshoot"
":POWer:SIGNals:SOURce:CURRent<i>"
":POWer:SIGNals:SOURce:VOLTage<i>"