Return Format
for <meas_spec> = DELay:
{ <edge_spec1> | <edge_spec2> | <edge_spec1>,<edge_spec2>} <NL>
for <meas_spec> = THResholds and <threshold mode> = PERCent:
THR,PERC,<upper>,<middle>,<lower><NL>
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
for <meas_spec> = THResholds and <threshold mode> = ABSolute:
THR,ABS,<upper>,<middle>,<lower><NL>
<upper>, <middle>, <lower> ::= A number specifying the upper, middle,
for <threshold spec> = STANdard:
THR,PERC,+90.0,+50.0,+10.0
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
"Introduction to :MEASure Commands"
":MEASure:DELay"
on page 430
":MEASure:SOURce"
":CHANnel<n>:RANGe"
":CHANnel<n>:SCALe"
":CHANnel<n>:PROBe"
":CHANnel<n>:UNITs"
and lower threshold percentage values
between Vbase and Vtop in NR3 format.
and lower threshold voltages in NR3
format.
on page 420
on page 458
on page 279
on page 280
on page 273
on page 281
21
:MEASure Commands
429