Compliance:usb:test:connection - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:COMPliance:USB:TEST:CONNection

Command Syntax
:COMPliance:USB:TEST:CONNection <connection>
<connection> ::= {SINGleended | DIFFerential}
When a Hi- Speed test has been selected, the
:COMPliance:USB:TEST:CONNection command specifies the test fixture
connection type:
• DIFFerential — specifies that a differential probe is used to probe the
• SINGleended — specifies that two single- ended SMA cables are used to
Query Syntax
:COMPliance:USB:TEST:CONNection?
The :COMPliance:USB:TEST:CONNection? query returns the specified the
test fixture connection type.
Return Format
<connection><NL>
<connection> ::= {SING | DIFF}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
signal under test.
In this case, use the :COMPliance:USB:SOURce:DIFFerential command to
specify the analog input channel connected to the differential probe.
probe the signal under test.
In this case, use the :COMPliance:USB:SOURce:DPLus command to
specify the analog input channel that is probing the D+ signal, and use
the :COMPliance:USB:SOURce:DMINus command to specify the analog
input channel that is probing the D- signal.
":COMPliance:USB:SOURce:DIFFerential"
":COMPliance:USB:SOURce:DPLus"
":COMPliance:USB:SOURce:DMINus"
:COMPliance Commands
on page 287
on page 289
on page 288
11