10
:CHANnel<n> Commands
:CHANnel<n>:PROBe:HEAD[:TYPE]
Command Syntax
This command is valid only for the 113xA Series probes.
N O T E
:CHANnel<n>:PROBe:HEAD[:TYPE] <head_param>
<head_param> ::= {SEND0 | SEND6 | SEND12 | SEND20 | DIFF0 | DIFF6
<n> ::= {1 | 2 | 3 | 4}
The :CHANnel<n>:PROBe:HEAD[:TYPE] command sets an analog channel
probe head type and dB value. You can choose from:
• SEND0 — Single- ended, 0dB.
• SEND6 — Single- ended, 6dB.
• SEND12 — Single- ended, 12dB.
• SEND20 — Single- ended, 20dB.
• DIFF0 — Differential, 0dB.
• DIFF6 — Differential, 6dB.
• DIFF12 — Differential, 12dB.
• DIFF20 — Differential, 20dB.
Query Syntax
:CHANnel<n>:PROBe:HEAD[:TYPE]?
The :CHANnel<n>:PROBe:HEAD[:TYPE]? query returns the current probe
head type setting for the selected channel.
Return Format
<head_param><NL>
<head_param> ::= {SEND0 | SEND6 | SEND12 | SEND20 | DIFF0 | DIFF6
See Also
274
(see
page
1208)
| DIFF12 | DIFF20 | NONE}