:CALibrate:OUTPut

Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | WAVEgen | WGEN1 | WGEN2}
Note: WAVE and WGEN1 are equivalent.
Note: WGEN2 only available on models with 2 WaveGen outputs.
The CALibrate:OUTPut command sets the signal that is available on the
rear panel TRIG OUT BNC:
• TRIGgers — pulse when a trigger event occurs.
• MASK — signal from mask test indicating a failure.
• WAVEgen, WGEN1, WGEN2 — waveform generator sync output signal.
Waveform
Type
SINusoid,
SQUare,
RAMP,
PULSe, SINC,
EXPRise,
EXPFall,
GAUSsian
DC, NOISe,
CARDiac
Query Syntax
:CALibrate:OUTPut?
The :CALibrate:OUTPut query returns the current source of the TRIG OUT
BNC signal.
Return Format
<signal><NL>
<signal> ::= {TRIG | MASK | WAVE | WGEN2}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
This signal depends on the :WGEN<w>:FUNCtion setting:
Sync Signal Characteristics
The Sync signal is a TTL positive pulse that occurs when the waveform rises
above zero volts (or the DC offset value).
N/A
"Introduction to :CALibrate Commands"
":WGEN<w>:FUNCtion"
on page 254
on page 1067
:CALibrate Commands
257