Agilent Technologies InfiniiVision 3000 User Manual

Agilent Technologies InfiniiVision 3000 User Manual

Infiniivision 3000 x-series
Hide thumbs Also See for InfiniiVision 3000:
Table of Contents

Advertisement

Agilent InfiniiVision
3000 X-Series
Oscilloscopes
User's Guide
s1

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies InfiniiVision 3000

  • Page 1 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 2: Safety Notices

    FAR 52.227-19 (June met. 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclo- sure of Software is subject to Agilent Tech- nologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will...
  • Page 3: Infiniivision 3000 X-Series Oscilloscopes-At A Glance

    DSO-X 3024A DSO-X 3034A DSO-X 3054A The Agilent InfiniiVision 3000 X- Series oscilloscopes deliver these features: • 100 MHz, 200 MHz, 350 MHz, and 500 MHz bandwidth models. • 2- and 4- channel digital storage oscilloscope (DSO) models. • 2+16- channel and 4+16- channel mixed- signal oscilloscope (MSO) models.
  • Page 4: Math Waveforms

    Quick Help. Complete instructions for using the quick help system are given in “Access the Built- In Quick Help" on page 45. For more information about InfiniiVision oscilloscopes, see: "www.agilent.com/find/scope" Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 5: Table Of Contents

    In This Guide This guide shows how to use the InfiniiVision 3000 X- Series oscilloscopes. When unpacking and using the • Chapter 1, “Getting Started,” starting on page 23 oscilloscope for the first time, see: When displaying waveforms and •...
  • Page 6 [Key1], then pressing Softkey2, then pressing Softkey3 are abbreviated as follows: Press [Key1]> Softkey2 > Softkey3. The keys may be a front panel [Key] or a Softkey. Softkeys are the six keys located directly below the oscilloscope display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 7 Use Auto Scale Compensate Passive Probes Learn the Front Panel Controls and Connectors Front Panel Overlays for Different Languages Learn the Rear Panel Connectors Learn the Oscilloscope Display Access the Built-In Quick Help Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 8 To adjust the vertical scale To adjust the vertical position To specify channel coupling To specify channel input impedance To specify bandwidth limiting To change the vertical scale knob's coarse/fine adjustment setting To invert a waveform Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 9 Math Transforms Differentiate Integrate FFT Measurement Square Root Ax + B Square Absolute Value Common Logarithm Natural Logarithm Exponential Base 10 Exponential Math Filters High Pass and Low Pass Filter Math Visualizations Magnify Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 10 To switch all digital channels on or off To switch groups of channels on or off To change the logic threshold for digital channels To reposition a digital channel To display digital channels as a bus Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 11 To assign a predefined label to a channel To define a new label To load a list of labels from a text file you create To reset the label library to the factory default Triggers Adjusting the Trigger Level Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 12 Trigger Mode/Coupling To select the Auto or Normal trigger mode To select the trigger coupling To enable or disable trigger noise rejection To enable or disable trigger HF Reject To set the trigger holdoff Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 13 High Resolution Acquisition Mode Acquiring to Segmented Memory Navigating Segments Measurements, Statistics, and Infinite Persistence with Segmented Memory Segmented Memory Re-Arm Time Saving Data from Segmented Memory Cursors To make cursor measurements Cursor Examples Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 14 Average DC RMS AC RMS Ratio Time Measurements Period Frequency Counter + Width – Width Burst Width Duty Cycle Rise Time Fall Time Delay Phase X at Min Y X at Max Y Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 15 Creating New Arbitrary Waveforms Editing Existing Arbitrary Waveforms Capturing Other Waveforms to the Arbitrary Waveform To output the waveform generator sync pulse To specify the expected output load To use waveform generator logic presets Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 16 To recall arbitrary waveforms Recalling Default Setups Performing a Secure Erase Print (Screens) To print the oscilloscope's display To set up network printer connections To specify the print options To specify the palette option Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 17 To check warranty and extended services status To contact Agilent To return the instrument Configuring the [Quick Action] Key Web Interface Accessing the Web Interface Browser Web Control Real Scope Remote Front Panel Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 18 Maximum input voltage at analog inputs Maximum input voltage at digital channels Environmental Conditions Probes and Accessories Passive Probes Single-Ended Active Probes Differential Probes Current Probes Accessories Available Loading Licenses and Displaying License Information Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 19: Utility Settings

    Interpreting CAN Lister Data Searching for CAN Data in the Lister Setup for LIN Signals LIN Triggering LIN Serial Decode Interpreting LIN Decode Interpreting LIN Lister Data Searching for LIN Data in the Lister Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 20 Setup for SPI Signals SPI Triggering SPI Serial Decode Interpreting SPI Decode Interpreting SPI Lister Data Searching for SPI Data in the Lister I2S Triggering and Serial Decode Setup for I2S Signals I2S Triggering Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 21 Searching for ARINC 429 Data in the Lister UART/RS232 Triggering and Serial Decode Setup for UART/RS232 Signals UART/RS232 Triggering UART/RS232 Serial Decode Interpreting UART/RS232 Decode UART/RS232 Totalizer Interpreting UART/RS232 Lister Data Searching for UART/RS232 Data in the Lister Index Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 22 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 23: Getting Started

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Getting Started Inspect the Package Contents Tilt the Oscilloscope for Easy Viewing Power-On the Oscilloscope Connect Probes to the Oscilloscope Input a Waveform Recall the Default Oscilloscope Setup Use Auto Scale Compensate Passive Probes...
  • Page 24 InfiniiVision 3000 X- Series oscilloscope. • Power cord (country of origin determines specific type). • Oscilloscope probes: • Two probes for 2- channel models. • Four probes for 4- channel models. • Documentation CD- ROM. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 25 N6450-61601 16-channel cable (qyt 1) 01650-82103 2-inch probe ground leads (qyt 5) 5090-4832 Grabber (qyt 20) Digital probe replacement parts are listed in the "Digital Channels" chapter. See Also • “Accessories Available" on page 322 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 26: Install The Optional Lan/Vga Or Gpib Module

    The LAN/VGA or GPIB module must be installed before powering on the oscilloscope. N O T E Tilt the Oscilloscope for Easy Viewing There are tabs under the oscilloscope's front feet that can be flipped out to tilt the oscilloscope. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 27: Power-On The Oscilloscope

    1 Connect the power cord to the rear of the oscilloscope, then to a oscilloscope suitable AC voltage source. Route the power cord so the oscilloscope's feet and legs do not pinch the cord. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 28: Connect Probes To The Oscilloscope

    The 50 input protection only functions when the oscilloscope is powered on. With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk With N2862A or N2863A 10:1 probe: 300 Vrms Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 29: Do Not Float The Oscilloscope Chassis

    Recall the Default Oscilloscope Setup To recall the default oscilloscope setup: 1 Press [Default Setup]. The default setup restores the oscilloscope's default settings. This places the oscilloscope in a known operating condition. The major default settings are: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 30: Use Auto Scale

    Use Auto Scale Use [Auto Scale] to automatically configure the oscilloscope to best display the input signals. 1 Press [Auto Scale]. You should see a waveform on the oscilloscope's display similar to this: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 31 25 Hz, a duty cycle greater than 0.5%, and an amplitude of at least 10 mV peak- to- peak. Any channels that do not meet these requirements are turned off. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 32: Compensate Passive Probes

    On the N2862/63/90 probes, the trimmer capacitor is the yellow adjustment on the probe tip. On other probes, the trimmer capacitor is located on the probe BNC connector. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 33: Learn The Front Panel Controls And Connectors

    The legend for these keys is directly above them, on the display. Their functions change as you navigate through the oscilloscope's menus. For the following figure, refer to the numbered descriptions in the table that follows. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 34 You can vary the intensity control to bring out signal detail, much like an analog oscilloscope. Digital channel waveform intensity is not adjustable. More details about using the Intensity control to view signal detail are on “To adjust waveform intensity" on page 129. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 35 16, “Waveform Generator,” starting on page 251. Trigger controls These controls determine how the oscilloscope triggers to capture data. Chapter 10, “Triggers,” starting on page 141 and Chapter 11, “Trigger Mode/Coupling,” starting on page 177. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 36: Horizontal Controls," Starting On

    For more information, see “Running, Stopping, and Making Single Acquisitions (Run Control)" on page 185. [Default Setup] Press this key to restore the oscilloscope's default settings (details on “Recall the Default Oscilloscope Setup" on page 29). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 37 • Multiplexed position knob — This position knob is used with Math, Ref, or Digital waveforms, whichever has the illuminated arrow to the left. For math and reference waveforms, the position knob acts like an analog channel vertical position knob. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 38 18, “Print (Screens),” starting on page 279. [Help] key Opens the Help Menu where you can display overview help topics and select the Language. See also “Access the Built-In Quick Help" page 45. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 39: Vertical Controls," Starting On

    61. Analog channel Attach oscilloscope probes or BNC cables to these BNC connectors. inputs With the InfiniiVision 3000 X-Series oscilloscopes, you can set the input impedance of the analog channels to either 50 or 1 M . See “To specify channel input impedance"...
  • Page 40: Digital Channels," Starting On

    To install a front panel overlay: 1 Gently pull on the front panel knobs to remove them. 2 Insert the overlay's side tabs into the slots on the front panel. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 41 4 Channel Overlay French 75019-94324 75019-94316 German 75019-94326 75019-94318 Italian 75019-94323 75019-94331 Japanese 75019-94311 75019-94312 Korean 75019-94329 75019-94321 Portuguese 75019-94327 75019-94319 Russian 75019-94322 75019-94315 Simplified Chinese 75019-94328 75019-94320 Spanish 75019-94325 75019-94317 Traditional Chinese 75019-94330 75019-94310 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 42: Learn The Rear Panel Connectors

    The video output connector is always active. For optimal video quality and performance, we recommend you use a shielded video cable with ferrite cores. A DSOXGPIB GPIB module may be ordered and installed separately. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 43: Learn The Oscilloscope Display

    PC to the oscilloscope via the USB device port. See “Remote Programming with Agilent IO Libraries" on page 307. Learn the Oscilloscope Display The oscilloscope display contains acquired waveforms, setup information, measurement results, and the softkey definitions. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 44 129. Information area The information area normally contains acquisition, analog channel, automatic measurement, and cursor results. Menu line This line normally contains menu name or other information associated with the selected menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 45: Access The Built-In Quick Help

    Press and hold front panel key or softkey (or right-click softkey when using web browser remote front panel). Quick Help remains on the screen until another key is pressed or a knob is turned. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 46 2 Repeatedly press and release the Language softkey or rotate the Entry language knob until the desired language is selected. The following languages are available: English, French, German, Italian, Japanese, Korean, Portuguese, Russian, Simplified Chinese, Spanish, and Traditional Chinese. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 47 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Horizontal Controls To adjust the horizontal (time/div) scale To adjust the horizontal delay (position) Panning and Zooming Single or Stopped Acquisitions To change the horizontal time mode (Normal, XY, or Roll) To display the zoomed time base...
  • Page 48: Horizontal Controls

    The current sample rate is displayed in the right- side information area. To adjust the horizontal (time/div) scale 1 Turn the large horizontal scale (sweep speed) knob marked change the horizontal time/div setting. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 49: To Adjust The Horizontal Delay (Position)

    The horizontal position knob works (in the Normal time mode) while acquisitions are running or when they are stopped. When running, adjusting the horizontal scale knob changes the sample rate. When Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 50: Panning And Zooming Single Or Stopped Acquisitions

    The screen will still contain a relatively good display if you zoom-in horizontally by a factor of 1000 and zoom-in vertically by a factor of 10 to display the information from where it was acquired. Remember that you can only make automatic measurements on displayed data. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 51: To Change The Horizontal Time Mode (Normal, Xy, Or Roll)

    To clear the display and restart an acquisition in Roll mode, press the [Single] key again. Use Roll mode on low- frequency waveforms to yield a display much like a strip chart recorder. It allows the waveform to roll across the display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 52: Xy Time Mode

    (assuming the amplitude is the same on both channels): sin = Signal must be centered in “ X” Measuring Signals 90 degrees Signals phase difference out of phase in phase Figure 3 XY time mode signals, centered on display Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 53 7 Calculate the phase difference using the formula below. For example, if the first Y value is 1.688 and the second Y value is 1.031: sin = second Y 1.031 first Y = 1.688 ; = 37.65 degrees of phase shift Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 54: To Display The Zoomed Time Base

    You can use Zoom to locate and horizontally expand part of the normal window for a more detailed (higher- resolution) analysis of signals. To turn on (or off) Zoom: 1 Press the zoom key (or press the [Horiz] key and then the Zoom softkey). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 55 The delay value, which is the time displayed relative to the trigger point) is momentarily displayed in the upper- right portion of the display when the delay time ( ) knob is turned. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 56: To Change The Horizontal Scale Knob's Coarse/Fine Adjustment Setting

    A small hollow triangle ( ) at the top of the display grid marks the position of the time reference. When delay time is set to zero, the trigger point indicator ( ) overlays the time reference indicator. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 57: Searching For Events

    3 Press Settings, and use the Search Settings Menu to set up the selected search type. Setting up searches is similar to setting up triggers: • For setting up Edge searches, see "Edge Trigger" on page 144. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 58: Triggers," Starting On

    Navigating the Time Base You can use the [Navigate] key and controls to navigate through: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 59: To Navigate Time

    • The Auto zoom softkey specifies whether the waveform display is automatically zoomed to fit the marked row as you navigate. • Pressing the Scroll Lister softkey lets you use the Entry knob to scroll through data rows in the Lister display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 60: To Navigate Segments

    In the Auto play mode: • Press the navigation keys to play backward, stop, or play forward in time. You can press the keys multiple times to speed up the playback. There are three speed levels. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 61 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Vertical Controls To turn waveforms on or off (channel or math) To adjust the vertical scale To adjust the vertical position To specify channel coupling To specify channel input impedance To specify bandwidth limiting...
  • Page 62: Vertical Controls

    1 and channel 2 are turned on and the menu for channel 2 is being displayed, to turn channel 1 off, press [1] to display the channel 1 menu; then, press [1] again to turn channel 1 off. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 63: To Adjust The Vertical Scale

    (see "To choose "expand about" center or ground" page 289). To specify channel coupling Coupling changes the channel's input coupling to either AC (alternating current) or DC (direct current). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 64: To Specify Channel Input Impedance

    50 Ohm — matches 50 ohm cables commonly used in making high • frequency measurements, and 50 ohm active probes. When 50 Ohm input impedance is selected, it is displayed with the channel information on- screen. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 65: To Specify Bandwidth Limiting

    Fine softkey in the Channel Menu) to toggle between fine and coarse adjustment of the vertical scale. When Fine adjustment is selected, you can change the channel's vertical sensitivity in smaller increments. The channel sensitivity remains fully calibrated when Fine is on. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 66: To Invert A Waveform

    Probe Menu. This menu lets you select additional probe parameters such as attenuation factor and units of measurement for the connected probe. The Channel Probe Menu changes depending on the type of probe connected. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 67: To Specify The Channel Units

    1 Press the channel key. 2 Press the Probe softkey until you have selected how you want to specify the attenuation factor, choosing either Ratio or Decibels. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 68: To Specify The Probe Skew

    200 ns difference. The skew setting is not affected by pressing [Default Setup] or [Auto Scale]. To calibrate a probe The Calibrate Probe softkey guides you through the process of calibrating probes. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 69 10:1 (+12 dB Atten) differential browser. • 10:1 (+20 dB Atten) single- ended browser. • 10:1 (+20 dB Atten) differential browser. 6 Press the Calibrate Probe softkey and follow the instructions on the display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 70 Vertical Controls For more information on InfiniiMax probes and accessories, see the probe's User's Guide. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 71: Math Waveforms

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Math Waveforms To display math waveforms To perform transforms or filters on an arithmetic operation To adjust the math waveform scale and offset Units for Math Waveforms Math Operators Math Transforms Math Filters Math Visualizations Math functions can be performed on analog channels.
  • Page 72 The vertical scaling and offset of each math function can be adjusted for ease of viewing and measurement considerations. The math function waveform can be measured using [Cursors] and/or [Meas]. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 73: To Perform Transforms Or Filters On An Arithmetic Operation

    If you manually set scale and offset for a function, select a new function, then select the original function, the original function will be automatically rescaled. See Also • "Units for Math Waveforms" on page 74 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 74: Units For Math Waveforms

    Math Operators Math operators perform arithmetic operations (like add, subtract, or multiply) on analog input channels. • "Add or Subtract" on page 75 • "Multiply or Divide" on page 75 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 75: Add Or Subtract

    74 Multiply or Divide When you select the multiply or divide math function, the Source 1 and Source 2 values are multiplied or divided point by point, and the result is displayed. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 76: Math Transforms

    Math transforms perform a transform function (like differentiate, integrate, FFT, or square root) on an analog input channel or on the result of an arithmetic operation. • "Differentiate" on page 77 • "Integrate" on page 78 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 77: Differentiate

    • d = differential waveform. • y = channel 1, 2, 3, or 4, or g(t) (internal arithmetic operation) data points. • i = data point index. • t = point- to- point time difference. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 78: Integrate

     dt plots the integral of the source using the "Trapezoidal Rule". The equation is: ∑ + Δ Where: • I = integrated waveform. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 79 Small DC offset in the integrate function input (or even small oscilloscope calibration errors) can cause the integrate function output to "ramp" up or down. This DC offset correction lets you level the integrate waveform. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 80 0 V level Integrate without DC offset correction Integrate with DC offset correction Figure 8 Integrate and Signal Offset See Also • "To perform transforms or filters on an arithmetic operation" page 73 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 81: Fft Measurement

    Press the Center softkey, then turn the Entry knob to set the desired center frequency of the display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 82 The FFT resolution is the quotient of the sampling rate and the number of FFT points (f /N). The current FFT Resolution is displayed above the softkeys. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 83 1. Set the horizontal scale to 50 µs/div, vertical sensitivity to 1 V/div, Units/div to 20 dBV, Offset to - 60.0 dBV, Center frequency to 250 kHz, frequency Span to 500 kHz, and window to Hanning. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 84 As frequency span is decreased, the number of points shown is reduced, and the display is magnified. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 85 • Adjust frequency span for better cursor placement. • Return to the Cursors Menu to fine tune the X cursor. For more information on the use of FFTs please refer to Agilent Application Note 243, The Fundamentals of Signal Analysis at "http://cp.literature.agilent.com/litweb/pdf/5952-...
  • Page 86 Aliasing happens when there are frequency components in the signal higher than half the sample rate. Because the FFT spectrum is limited by this frequency, any higher components are displayed at a lower (aliased) frequency. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 87 In order to minimize spectral leakage, windows that approach zero smoothly at the beginning and end of the signal are employed as filters to the FFT. The FFT Menu provides four windows: Hanning, Flat Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 88: Square Root

    Math Waveforms Top, Rectangular, and Blackman- Harris. For more information on leakage, see Agilent Application Note 243, The Fundamentals of Signal Analysis at "http://cp.literature.agilent.com/litweb/pdf/5952- 8898E.pdf." Square Root Square root ( ) calculates the square root of the selected source. Where the transform is undefined for a particular input, holes (zero values) appear in the function output.
  • Page 89: Square

    Press the Source softkey to select the signal source. See Also • "Square Root" on page 88 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 90: Absolute Value

    DSOX3ADVMATH advanced math measurements license) performs a transform of the input source. Where the transform is undefined for a particular input, holes (zero values) appear in the function output. See Also • "Natural Logarithm" on page 91 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 91: Natural Logarithm

    "Common Logarithm" on page 90 Exponential The Exponential (e^x) function (available with the DSOX3ADVMATH advanced math measurements license) performs a transform of the input source. See Also • "Base 10 Exponential" on page 92 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 92: Base 10 Exponential

    • "High Pass and Low Pass Filter" on page 93 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 93: High Pass And Low Pass Filter

    The ratio of the input signal's Nyquist frequency and the selected -3 dB cutoff frequency N O T E affects how many points are available in the output, and under some circumstances, there are no points in the output waveform. Figure 15 Example of Low Pass Filter Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 94: Math Visualizations

    Figure 16 Example of Magnify See Also • "Ax + B" on page 88 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 95: Measurement Trend

    Use the Type: softkey to select the measurement whose trend you want to look at. You can display trend values for these measurements: • Average • RMS - AC • Ratio • Period • Frequency • +Width • - Width Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 96: Chart Logic Bus Timing

    Figure 18 Example of Chart Logic Bus Timing Use the Units/Code softkey to specify the analog value equivalent of each increment in the bus data value. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 97: Chart Logic Bus State

    Use the More Chart softkey to open a submenu for specifying the analog value equivalent of each bus value increment, the analog equivalent of a zero bus value, and the type of values the charted bus data represents (volts, amps, etc.). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 98 Use the 0 Offset softkey to specify the analog value equivalent of a bus data value of zero. Use the Units softkey to specify the type of values the bus data represents (volts, amps, etc.). See Also • "Chart Logic Bus Timing" on page 96 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 99: Reference Waveforms

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Reference Waveforms To save a waveform to a reference waveform location To display a reference waveform To scale and position reference waveforms To adjust reference waveform skew To display reference waveform information To save/recall reference waveform files to/from a USB storage...
  • Page 100: To Display A Reference Waveform

    2 In the Reference Waveform Menu, press the Ref softkey and turn the Entry knob to select the desired reference waveform location. 3 Then, press the Ref softkey again to enable/disable the reference waveform display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 101: To Scale And Position Reference Waveforms

    2 Turn the upper multiplexed knob to adjust the reference waveform scale. 3 Turn the lower multiplexed knob to adjust the reference waveform position. To adjust reference waveform skew Once reference waveforms are displayed, you can adjust their skew. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 102: To Display Reference Waveform Information

    You can recall a reference waveform file from a USB storage device into one of the reference waveform locations. See "To recall reference waveform files from a USB storage device" on page 275. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 103: Digital Channels

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Digital Channels To connect the digital probes to the device under test Acquiring waveforms using the digital channels To display digital channels using AutoScale Interpreting the digital waveform display To switch all digital channels on or off...
  • Page 104: Probe Cable For Digital Channels

    You do not need to power- off the oscilloscope. Probe cable for digital channels C A U T I O N Use only the Agilent logic probe and accessory kit supplied with the mixed-signal oscilloscope (see "Accessories Available" on page 322).
  • Page 105 4 Connect a grabber to one of the probe leads. (Other probe leads are omitted from the figure for clarity.) Grabber 5 Connect the grabber to a node in the circuit you want to test. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 106 Signal Ground Grabber 7 Repeat these steps until you have connected all points of interest. Signals Ground Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 107: Acquiring Waveforms Using The Digital Channels

    When signals are connected to the digital channels — be sure to connect the ground leads — AutoScale quickly configures and displays the digital channels. • To configure the instrument quickly, press the [AutoScale] key. Figure 20 Example: AutoScale of digital channels (MSO models only) Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 108: Interpreting The Digital Waveform Display

    "How AutoScale Works" page 31. To set the instrument to the factory- default configuration, press the [Default Setup] key. Interpreting the digital waveform display The following figure shows a typical display with digital channels. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 109: To Change The Displayed Size Of The Digital Channels

    To change the displayed size of the digital channels 1 Press the [Digital] key. 2 Press the size ( ) softkey to select how the digital channels are displayed. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 110: To Switch A Single Channel On Or Off

    Each time you press the softkey, the softkey's mode toggles between Turn on and Turn off. To change the logic threshold for digital channels 1 Press the [Digital] key so that the Digital Channel Menu is displayed. 2 Press the Thresholds softkey Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 111: To Reposition A Digital Channel

    Dnn designation (where nn is a one or two digit channel number) to D*. The "*" indicates that two channels are overlaid. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 112: To Display Digital Channels As A Bus

    "Using cursors to read bus values" on page 113). The Base softkey lets you choose to display the bus values in hex or binary. The buses are shown at the bottom of the display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 113 3 Press the Source softkey and select Bus1 or Bus2. 4 Use the Entry knob and the X1 and X2 softkeys to position the cursors where you want to read the bus values. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 114 — rising edge ( ) or falling edge ( ) — are included in the pattern specification. A byte that consists of all don't cares (X) will be displayed in the bus as a don't care (X). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 115: Digital Channel Signal Fidelity: Probe Impedance And Grounding

    The design of the probes minimizes the first problem, while the second is easily addressed by good probing practices. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 116: Input Impedance

    This high- frequency model takes into account pure tip capacitance to ground as well as series tip resistance, and the cable's characteristic impedance (Z Figure 21 DC and Low-Frequency Probe Equivalent Circuit Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 117 The stray tip capacitance, which is generally small (1 pF), sets the final break point on the impedance chart. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 118: Probe Grounding

    Increased length in this path will, at high frequencies, create large common mode voltages at the probe input. The voltage generated behaves as if this path were an inductor according to the equation: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 119 LC circuit at the input of the probe, is also increased. Because the digital channels display reconstructed waveforms, they do not show ringing and perturbations. You will not find ground problems through examination of the waveform display. In fact, it is likely Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 120: Best Probing Practices

    If you need to remove a probe lead from the cable, insert a paper clip or other small pointed object into the side of the cable assembly, and push to release the latch while pulling out the probe lead. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 121 Replacement 2-inch probe grounds (qty 5) 5959-9335 Replacement pod ground leads (qty 5) 5090-4833 Grabbers (qty 20) 01650-94309 Package of probe labels For other replacement parts, consult the InfiniiVision 2000/3000 X- Series Oscilloscopes Service Guide. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 122 Digital Channels Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 123: Serial Decode

    The threshold level for each serial signal can be set in the Signals Menu. Press the [Serial] key, then the Signals softkey. Serial Decode Options Agilent's hardware- accelerated serial decode options can be installed when the oscilloscope is manufactured, or added later. The following serial decode licenses are available: •...
  • Page 124: Lister

    296. To order serial decode licenses, go to "www.agilent.com" and search for the product number (for example, DSOX3AUTO) or contact your local Agilent Technologies representative (see "www.agilent.com/find/contactus)." Lister Lister is a powerful tool for investigating protocol failures. You can use Lister to view large amounts of packet level serial data in a tabular format, including time tags and specific decoded values.
  • Page 125 6 Press the Zoom to Selection softkey (or push the Entry knob) to center the waveform display at the time associated with the selected Lister row and automatically set the horizontal scale setting. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 126: Searching Lister Data

    The Search softkey lets you specify events to find. It is similar to specifying protocol triggers. Events found are marked in orange in the far left Lister column. The total number of events found is displayed above the softkeys. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 127 "Searching for MIL- STD- 1553 Data in the Lister" on page 397 • "Searching for SPI Data in the Lister" on page 378 • "Searching for UART/RS232 Data in the Lister" on page 415 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 128 Serial Decode Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 129: Display Settings

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Display Settings To adjust waveform intensity To set or clear persistence To clear the display To select the grid type To adjust the grid intensity To freeze the display To adjust waveform intensity You can adjust the intensity of displayed waveforms to account for various signal characteristics, such as fast time/div settings and low trigger rates.
  • Page 130 Display Settings Figure 25 Amplitude Modulation Shown at 100% Intensity Figure 26 Amplitude Modulation Shown at 40% Intensity Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 131: To Set Or Clear Persistence

    When variable persistence is selected, press the Time softkey and use the Entry knob to specify the amount of time that previous acquisitions are to be displayed. The display will begin accumulating multiple acquisitions. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 132: To Clear The Display

    IRE units.) The mV and IRE grid values are accurate (and match Y cursor values) when the vertical scaling is 140 mV/division and the vertical offset is 245 mV. To select the grid type: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 133: To Adjust The Grid Intensity

    2 To un- freeze the display, press [Quick Action] again. Manual cursors can be used on the frozen display. Many activities, such as adjusting the trigger level, adjusting vertical or horizontal settings, or saving data will un- freeze the display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 134 Display Settings Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 135: Labels

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Labels To turn the label display on or off To assign a predefined label to a channel To define a new label To load a list of labels from a text file you create...
  • Page 136: To Assign A Predefined Label To A Channel

    To assign a predefined label to a channel 1 Press the [Label] key. 2 Press the Channel softkey, then turn the Entry knob or successively press the Channel softkey to select a channel for label assignment. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 137: To Define A New Label

    The channel does not have to be turned on to have a label assigned to it. If the channel is turned on, its current label will be highlighted. 3 Press the Spell softkey; then, turn the Entry knob to select the first character in the new label. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 138: To Load A List Of Labels From A Text File You Create

    1 Use a text editor to create each label. Each label can be up to ten characters in length. Separate each label with a line feed. 2 Name the file labellist.txt and save it on a USB mass storage device such as a thumb drive. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 139: To Reset The Label Library To The Factory Default

    Defaulting labels without erasing the default library N O T E Pressing [Default Setup] sets all channel labels back to the default labels but does not erase the list of user-defined labels in the library. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 140 Labels Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 141: Triggers

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Triggers Adjusting the Trigger Level Forcing a Trigger Edge Trigger Edge then Edge Trigger Pulse Width Trigger Pattern Trigger OR Trigger Rise/Fall Time Trigger Nth Edge Burst Trigger Runt Trigger Setup and Hold Trigger...
  • Page 142 (horizontal position) setting (see "To adjust the horizontal delay (position)" on page 49). Trigger Event Post-Trigger Buffer Pre-Trigger Buffer Acquisition Memory Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 143: Adjusting The Trigger Level

    Trigger] to force a trigger and see what the input signals look like. In the Auto trigger mode, when the trigger condition is not met, triggers are forced and the "Auto?" indicator is displayed. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 144: Edge Trigger

    4 Press the Slope softkey and select rising edge, falling edge, alternating edges, or either edge (depending on the selected source). The selected slope is displayed in the upper- right corner of the display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 145 You can then pan and zoom through the data using the Horizontal and Vertical knobs to find a stable trigger point. AutoScale often produces a triggered display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 146: Edge Then Edge Trigger

    1 Press the [Trigger] key. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select Edge then Edge. 3 Press the Sources softkey. 4 In the Edge then edge Sources Menu: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 147: Pulse Width Trigger

    Pattern trigger in the Trigger Menu (see "Pattern Trigger" on page 150). 1 Press the [Trigger] key. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select Pulse Width. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 148 The selected pulse polarity is displayed in the upper- right corner of the display. A positive pulse is higher than the current trigger level or threshold and a negative pulse is lower than the current trigger level or threshold. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 149 2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz bandwidth models). • 10 ns to 10 s for >< qualifier, with minimum difference of 5 ns between upper and lower settings. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 150: Pattern Trigger

    In this case, the trigger occurs when the > softkey time value is exceeded (not when the pattern exits). >< (In Range) – when the pattern is present for a time within a range • of values. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 151 5 For each channel you select, press the Pattern softkey; then, turn the Entry knob to set the condition for that channel in the pattern. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 152 If no edge is specified, the oscilloscope will trigger on the last edge that makes the pattern true. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 153: Hex Bus Pattern Trigger

    "Bus values are displayed when using Pattern trigger" on page 114. OR Trigger The OR trigger mode triggers when any one (or more) of the specified edges on analog or digital channels is found. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 154 5 For each channel you select, press the Slope softkey and select (Rising), (Falling), (Either), or X (Don't Care). The selected slope is displayed above the softkeys. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 155: Rise/Fall Time Trigger

    High level Low level Rising edge time Falling edge time 1 Press the [Trigger] key. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select Rise/Fall Time. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 156: Nth Edge Burst Trigger

    8 Press the Time softkey, and turn the Entry knob to select the time. Nth Edge Burst Trigger The Nth Edge Burst trigger lets you trigger on the Nth edge of a burst that occurs after an idle time. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 157 4 Press the Slope softkey to specify the slope of the edge. 5 Press the Idle softkey; then, turn the Entry knob to specify the idle time. 6 Press the Edge softkey; then, turn the Entry knob to which edge number to trigger on. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 158: Runt Trigger

    • A negative runt pulse crosses through an upper threshold but not a lower threshold. To trigger on runt pulses: 1 Press the [Trigger] key. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select Runt. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 159 This lets you specify that a runt pulse be less than or greater than a certain width. 8 If you selected the "less than" or "greater than" Qualifier, press the Time softkey; then, turn the Entry knob to select the time. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 160: Setup And Hold Trigger

    7 Set the appropriate trigger tevel for the data signal using the Trigger Level knob. 8 Press the < Setup softkey, and turn the Entry knob to select the setup time. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 161: Video Trigger

    It is important, when using a 10:1 passive probe, that the probe is correctly compensated. N O T E The oscilloscope is sensitive to this and will not trigger if the probe is not properly compensated, especially for progressive formats. 1 Press the [Trigger] key. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 162 Many video signals are produced from 75 sources. To provide correct matching to these sources, a 75 terminator (such as an Agilent 11094B) should be connected to the oscilloscope input. 4 Press the sync polarity softkey to set the Video trigger to either positive ) or negative ( ) sync polarity.
  • Page 163 Tri-level HDTV 720p/60 Progressive Tri-level HDTV 1080p/24 Progressive Tri-level HDTV 1080p/25 Progressive Tri-level HDTV 1080p/30 Progressive Tri-level HDTV 1080p/50 Progressive Tri-level HDTV 1080p/60 Progressive Tri-level HDTV 1080i/50 Interlaced Tri-level HDTV 1080i/60 Interlaced Tri-level Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 164 Line: Field1 and Line:Field2 — Trigger on the selected line # in field 1 or • field 2 (interlaced standards only). Line: Alternate — Alternately trigger on the selected line # in field 1 • and field 2 (NTSC, PAL, PAL- M, and SECAM only). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 165 "To trigger on a specific field of the video signal" on page 168 • "To trigger on all fields of the video signal" on page 169 • "To trigger on odd or even fields" on page 170 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 166: To Set Up Generic Video Triggers

    Turning the trigger Level knob in Video trigger does not change the trigger level because the trigger level is automatically set to the sync pulse tips. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 167 Field 1 and Field 2. This is a quick way to compare the Field 1 VITS and Field 2 VITS or to check for the correct insertion of the half line at the end of Field 1. Figure 27 Example: Triggering on Line 136 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 168: To Trigger On All Sync Pulses

    Field 2 (available for interleaved standards). When a specific field is selected, the oscilloscope triggers on the rising edge of the first serration pulse in the vertical sync interval in the specified field (1 or 2). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 169: To Trigger On All Fields Of The Video Signal

    2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select Video. 3 Press the Settings softkey, then press the Standard softkey to select the appropriate TV standard. 4 Press the Mode softkey and select All Fields. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 170: To Trigger On Odd Or Even Fields

    But this definition of field does not take into consideration the phase of the reference subcarrier. When Field 1 is selected, the trigger system will find any field where the vertical sync Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 171 Repeat until the correct phase is displayed. When holdoff is adjusted using the Field Holdoff softkey and the Entry knob, the corresponding holdoff time will be displayed in the Trigger Mode and Coupling Menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 172 10 ms HDTV 720p/60 8.35 ms HDTV 1080p/24 20.835 ms HDTV 1080p/25 20 ms HDTV 1080p/30 20 ms HDTV 1080p/50 16.67 ms HDTV 1080p/60 8.36 ms HDTV 1080i/50 10 ms HDTV 1080i/60 8.35 ms Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 173: Usb Trigger

    1 Press [Default Setup]. 2 Press the [Label] key to switch labels on. 3 Turn on any analog or digital channels that you will be using for the USB signals. 4 Press the [Trigger] key. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 174 You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s). 8 Press the D+ and D– softkeys to select the channel connected to the USB signal D+ and D– lines. The D+ and D– labels for the source channels are automatically set. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 175: Serial Trigger

    362 • "I2S Triggering" on page 384 • "LIN Triggering" on page 345 • "MIL- STD- 1553 Triggering" on page 393 • "SPI Triggering" on page 373 • "UART/RS232 Triggering" on page 409 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 176 Triggers Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 177: Trigger Mode/Coupling

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Trigger Mode/Coupling To select the Auto or Normal trigger mode To select the trigger coupling To enable or disable trigger noise rejection To enable or disable trigger HF Reject To set the trigger holdoff...
  • Page 178: To Select The Auto Or Normal Trigger Mode

    This may be more likely, for example, when the horizontal scale knob is set to a slow time/div setting, such as 500 ms/div. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 179 Trig'd? to flash (this tells you the pre- trigger buffer is filled). See Also • "Forcing a Trigger" on page 143 • "To set the trigger holdoff" on page 182 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 180: To Select The Trigger Coupling

    Use LF Reject coupling to get a stable edge trigger when your waveform has low frequency noise. TV coupling — is normally grayed- out, but is automatically selected • when TV trigger is enabled in the Trigger Menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 181: To Enable Or Disable Trigger Noise Rejection

    1 Press the [Mode/Coupling] key. 2 In the Trigger Mode and Coupling Menu, press the HF Reject softkey to enable or disable. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 182: To Set The Trigger Holdoff

    Changing the time base settings does not affect the trigger holdoff time. With Agilent's MegaZoom technology, you can press [Stop], then pan and zoom through the data to find where the waveform repeats. Measure this time using cursors;...
  • Page 183: Maximum Voltage At Oscilloscope External Trigger Input

    4 Press the Probe softkey; then, turn the entry knob to specify the probe attenuation. The attenuation factor can be set from 0.1:1 to 1000:1 in a 1- 2- 5 sequence. The probe attenuation factor must be set properly for measurements to be made correctly. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 184 Trigger Mode/Coupling Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 185: Acquisition Control

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Acquisition Control Running, Stopping, and Making Single Acquisitions (Run Control) Overview of Sampling Selecting the Acquisition Mode Acquiring to Segmented Memory This chapter shows how to use the oscilloscope's acquisition and run controls.
  • Page 186 To acquire data with the longest possible record length, press the [Single] key. For more information on settings that affect record length, see "Length Control" on page 269. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 187: Overview Of Sampling

    Aliasing Aliasing occurs when signals are under- sampled (f < 2f ). Aliasing is the signal distortion caused by low frequencies falsely reconstructed from an insufficient number of sample points. Figure 33 Aliasing Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 188: Oscilloscope Bandwidth And Sample Rate

    (square waves are made up of sine waves at the fundamental frequency and an infinite number of odd harmonics), and typically, for 1 Ghz bandwidths and below, oscilloscopes have a Gaussian frequency response. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 189: Oscilloscope Rise Time

    . This way, there is less aliasing, and aliased frequency components have a greater amount of attenuation. See Also Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity: How to Make the Most Accurate Digital Measurements, Agilent Application Note 1587 ("http://cp.literature.agilent.com/litweb/pdf/5989- 5732EN.pdf")
  • Page 190: Oscilloscope Bandwidth Required

    = 1.0 x f knee = 1.3 x f knee = 1.9 x f knee See Also Choosing an Oscilloscope with the Right Bandwidth for your Application, Agilent Application Note 1588 ("http://cp.literature.agilent.com/litweb/pdf/5989- 5733EN.pdf") Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 191: Memory Depth And Sample Rate

    1 Press the [Acquire] key on the front panel. 2 In the Acquire Menu, press the Acq Mode softkey; then, turn the Entry knob to select the acquisition mode. The InfiniiVision oscilloscopes have the following acquisition modes: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 192: Normal Acquisition Mode

    This mode displays all pulses that are at least as wide as the sample period. For InfiniiVision 3000 X- Series oscilloscopes, which have a maximum sample rate of 4 GSa/s, a sample is taken every 250 ps (sample period).
  • Page 193 Normal acquire mode, making them easier to see. To characterize the glitch, use the cursors or the automatic measurement capabilities of the oscilloscope. Figure 36 Sine With Glitch, Normal Mode Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 194 4 Characterize the glitch with Zoom mode: a Press the zoom key (or press the [Horiz] key and then the Zoom softkey). b To obtain a better resolution of the glitch, expand the time base. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 195: Averaging Acquisition Mode

    2 Press the #Avgs softkey and turn the Entry knob to set the number of averages that best eliminates the noise from the displayed waveform. The number of acquisitions being averaged is displayed in the # Avgs softkey. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 196 Acquisition Control Figure 38 Random noise on the displayed waveform Figure 39 128 Averages used to reduce random noise Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 197: High Resolution Acquisition Mode

    Acquiring to Segmented Memory You can purchase the oscilloscope with the segmented memory option factory- installed (Option SGM) or you can install a license to enable it (order model number DSOX3SGM "Segmented Memory"). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 198 If the signal you are measuring has more than about 1 s of inactivity, consider selecting the Normal trigger mode to prevent AutoTriggering. See "To select the Auto or Normal trigger mode" on page 178. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 199: Navigating Segments

    To perform measurements and view statistical information, press [Meas] and set up your desired measurements (see Chapter 14, “Measurements,” starting on page 213). Then, press Analyze Segments. Statistical data will be accumulated for the measurements you have chosen. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 200: Segmented Memory Re-Arm Time

    Saving Data from Segmented Memory You can save either the currently displayed segment (Save Segment - Current), or all segments (Save Segment - All) in the following data formats: CSV, ASCII XY, and BIN. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 201 When the oscilloscope is busy saving multiple segments, progress is displayed in the upper right area of the display. For more information, see "To save CSV, ASCII XY, or BIN data files" page 266. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 202 Acquisition Control Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 203 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Cursors To make cursor measurements Cursor Examples Cursors are horizontal and vertical markers that indicate X- axis values and Y- axis values on a selected waveform source. You can use cursors to make custom voltage, time, phase, or ratio measurements on oscilloscope signals.
  • Page 204: To Make Cursor Measurements

    3 In the Cursors Menu, press Mode; then, select the desired mode: Manual — X, 1/ X, and Y values are displayed. X is the difference • between the X1 and X2 cursors and Y is the difference between the Y1 and Y2 cursors. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 205 5 Select the cursor(s) to be adjusted: • Push the Cursors knob; then, turn the Cursors knob. To finalize your selection, either push the Cursors knob again or wait about five seconds for the popup menu to disappear. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 206 For phase or ratio units, once the 0 and 360 degree or 0 and 100% locations are set, adjusting cursors will display measurements relative to the set locations. 7 Adjust the selected cursor(s) by turning the Cursors knob. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 207: Cursor Examples

    Cursors Cursor Examples Figure 40 Cursors used to measure pulse widths other than middle threshold points Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 208 Cursors Figure 41 Cursors measure frequency of pulse ringing Expand the display with Zoom mode, then characterize the event of interest with the cursors. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 209 Cursors Figure 42 Cursors track Zoom window Put the X1 cursor on one side of a pulse and the X2 cursor on the other side of the pulse. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 210 Cursors Figure 43 Measuring pulse width with cursors Press the X1 X2 linked softkey and move the cursors together to check for pulse width variations in a pulse train. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 211 Cursors Figure 44 Moving the cursors together to check pulse width variations Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 212 Cursors Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 213: Measurements

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Measurements To make automatic measurements Measurements Summary Voltage Measurements Time Measurements Count Measurements Mixed Measurements Measurement Thresholds Measurement Window with Zoom Display Measurement Statistics The [Meas] key lets you make automatic measurements on waveforms.
  • Page 214: To Make Automatic Measurements

    "< value", or "> value", or a similar message to indicate that the measurement may not be reliable. 3 Press the Type: softkey; then, rotate the Entry knob to select a measurement to be made. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 215: Measurements Summary

    All measurements except Counter are available for reference waveforms and math waveforms other than FFT. A limited set of measurements is available for math FFT waveforms and for digital channel waveforms (as described in the following table). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 216 219 "Rising Edge Count" page 233 "Falling Edges Count" page 233 "Positive Pulse Count" page 232 "Negative Pulse Count" page 233 "Overshoot" on page 221 "Peak-Peak" on page 219 "Period" on page 226 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 217: Snapshot All

    Power Measurement Application User's Guide at "www.agilent.com/find/3000X- Series- manual" or on the Documentation CD. Snapshot All The Snapshot All measurement type displays a popup containing a snapshot of all the single waveform measurements. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 218: Voltage Measurements

    Probe Units softkey. See "To specify the channel units" page 67. The units of math waveforms are described in "Units for Math Waveforms" on page 74. • "Peak- Peak" on page 219 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 219: Peak-Peak

    The Top of a waveform is the mode (most common value) of the upper part of the waveform, or if the mode is not well defined, the top is the same as Maximum. The Y cursor shows the value being measured. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 220: Base

    The Base of a waveform is the mode (most common value) of the lower part of the waveform, or if the mode is not well defined, the base is the same as Minimum. The Y cursor shows the value being measured. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 221: Overshoot

    Maximum − D Top Rising edge overshοot = × 100 Amplitude Falling edge overshοot = Base − D local Minimum × 100 Amplitude Overshoot local Maximum Base local Minimum Overshoot Figure 46 Automatic Overshoot measurement Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 222: Preshoot

    The N Cycles measurement interval variation measures the value on an integral number of periods of the displayed signal. If less than three edges are present, the measurement shows "No edges". The X cursors show what interval of the waveform is being measured. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 223: Dc Rms

    The standard deviation of a measurement is the amount that a measurement varies from the mean value. The Mean value of a measurement is the statistical average of the measurement. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 224 • N = the number of measurements taken. • x = the ith measurement result. The standard deviation is calculated as follows: ∑ − ¯ where: • = the standard deviation. • N = the number of measurements taken. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 225: Ratio

    "– Width" on page 228 • "Burst Width" on page 228 • "Duty Cycle" on page 228 • "Rise Time" on page 229 • "Fall Time" on page 229 • "Delay" on page 229 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 226: Period

    You may need to change the measurement window setting so that the measurement is made in the lower, Zoom window. See "Measurement Window with Zoom Display" on page 236. If the waveform is clipped, it may not be possible to make the measurement. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 227: Counter

    Isolating an event for Frequency measurement Counter The InfiniiVision 3000 X- Series oscilloscopes have an integrated hardware frequency counter which counts the number of cycles that occur within a period of time (known as the gate time) to measure the frequency of a signal.
  • Page 228: Width

    The X cursors show the time period being measured. The Y cursor shows the middle threshold point. Duty cycle = + Width Period × 100 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 229: Rise Time

    4 Press the Settings softkey to select the second analog channel source and slope for the delay measurement. The default Delay settings measure from the rising edge of channel 1 to the rising edge of channel 2. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 230: Phase

    Negative phase shift values indicate that the rising edge of source 1 occurred after the rising edge of source 2. Delay Phase = Source 1 Period × 360 Period Source 1 Delay Source 2 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 231: At Min Y

    Minimum, starting from the left- side of the display. For periodic signals, the position of the minimum may vary throughout the waveform. The X cursor shows where the current X at Min Y value is being measured. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 232: At Max Y

    "Rising Edge Count" on page 233 • "Falling Edges Count" on page 233 Positive Pulse Count The Positive Pulse Count measurement is a pulse count for the selected waveform source. Positive pulse count Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 233: Negative Pulse Count

    This measurement is available for analog channels. Mixed Measurements • "Area" on page 233 Area Area measures the area between the waveform and the ground level. Area below the ground level is subtracted from area above the ground level. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 234: Measurement Thresholds

    You can also open the Measurement Threshold Menu by pressing [Analyze] > Features and then selecting Measurement Thresholds. 2 Press the Source softkey to select the analog channel or math waveform source for which you want to change measurement thresholds. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 235 The middle value is bounded by the values set for lower and upper thresholds. The default middle threshold is 50% or 1.20 V. If threshold Type is set to %, the middle threshold value can be set • from 6% to 94%. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 236: Measurement Window With Zoom Display

    1. Select the measurements you desire for the channels you are using (see "Measurements Summary" on page 215). From the Measurement Menu, press the Statistics softkey to enter the Statistics Menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 237 When you navigate away from the Measurements Menu, the statistics display will no longer be displayed, but statistical data collection will continue. Return to the Measurements Menu to see the data again. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 238 Segments softkey to accumulate measurement statistics for the acquired segments. You can also turn on infinite persistence (in the Display Menu) and press the Analyze Segments softkey to create an infinite persistence display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 239: Mask Testing

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Mask Testing To create a mask from a "golden" waveform (Automask) Mask Test Setup Options Mask Statistics To manually modify a mask file Building a Mask File One way to verify a waveform's compliance to a particular set of parameters is to use mask testing.
  • Page 240 (volts or seconds), selectable using the Units softkey. 8 Press the Create Mask softkey. The mask is created and testing begins. Whenever the Create Mask softkey is pressed the old mask is erased and a new mask is created. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 241: Mask Test Setup Options

    ± Y and ± X settings. They may be set so large that the mask is not visible. Mask Test Setup Options From the Mask Test Menu, press the Setup softkey to enter the Mask Setup Menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 242 (which is tied to the amount of failures per test). The sigma value may exceed the selected value when a small sigma value is chosen. The actual sigma is displayed. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 243 The Source softkey in the Mask Setup Menu is the same as the Source softkey in the Automask Menu. Test All When enabled, all displayed analog channels are included in the mask test. When disabled, just the selected source channel is included in the test. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 244: Mask Statistics

    • Number of failures (acquisitions in which the signal excursion went beyond the mask). • Failure rate (percentage of failures). • Sigma (the ratio of process sigma to maximum achievable sigma, based on number of waveforms tested). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 245: To Manually Modify A Mask File

    9 Open the .msk file your created using a text editor (such as Wordpad). 10 Edit, save, and close the file. The mask file contains the following sections: • Mask File Identifier. • Mask Title. • Mask Violation Regions. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 246 "MIN" for the first and last points. The top region must be the lowest numbered region in the file. The bottom region must be the highest numbered region in the file. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 247 The oscilloscope setup information contains remote programming language commands that the oscilloscope executes when the mask file is loaded. Any legal remote programming command can be entered in this section. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 248: Building A Mask File

    X and Y values in the mask file. The following mask file produced the mask shown above: MASK_FILE_548XX "All Regions" /* Region Number */ 1 /* Number of vertices */ 4 -12.50, -10.00, 1.750 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 249: Mask Testing

    /* Number of vertices */ 5 10.00, -1.000 12.50, -0.500 15.00, -0.500 15.00, -1.500 12.50, -1.500 /* Region Number */ 8 /* Number of vertices */ 4 -12.50, -10.00, -1.750 10.00, -1.750 12.50, setup :MTES:ENAB 1 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 250: How Is Mask Testing Done

    The number of acquisitions tested is also logged and displayed as "# of Tests". The mask file allows greater resolution than the 200 X 640 database. Some quantization of data occurs to reduce the mask file data for display on- screen. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 251: To Select Generated Waveform Types And Settings

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Waveform Generator To select generated waveform types and settings To edit arbitrary waveforms To output the waveform generator sync pulse To specify the expected output load To use waveform generator logic presets To add noise to the waveform generator output To restore waveform generator defaults A waveform generator is built into the oscilloscope.
  • Page 252: Waveform Generator

    "To edit arbitrary waveforms" on page 255. Sine Use the Frequency/Frequency 100 mHz to 20 mVpp to ±2.50 V Fine/Period/Period Fine, Amplitude/High-Level, 20 MHz 5 Vpp and Offset/Low-Level softkeys to set the sine signal parameters. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 253 Exponential Use the Frequency/Frequency 100 mHz to 20 mVpp to ±2.50 V Fall Fine/Period/Period Fine, Amplitude/High-Level, 5 MHz 5 Vpp and Offset/Low-Level softkeys to set the exponential fall signal parameters. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 254 259 • "To use waveform generator logic presets" on page 260 • "To add noise to the waveform generator output" on page 261 • "To restore waveform generator defaults" on page 261 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 255: To Edit Arbitrary Waveforms

    Softkey Description Create New Opens the New Waveform Menu. See "Creating New Arbitrary Waveforms" on page 256. Edit Existing Opens the Edit Waveform Points Menu. See "Editing Existing Arbitrary Waveforms" page 257. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 256: Creating New Arbitrary Waveforms

    3 Use the Amplitude/High-Level and Offset/Low-Level softkeys to set the voltage bounding parameter of the arbitrary waveform. 4 When you are ready to create the new arbitrary waveform, press Apply & Edit. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 257: Editing Existing Arbitrary Waveforms

    The Edit Waveform Points Menu is opened by pressing Edit Existing in the Edit Waveform Menu or by pressing Apply & Edit when creating a new arbitrary waveform. To specify the voltage values of points: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 258: Capturing Other Waveforms To The Arbitrary Waveform

    To capture another waveform to the arbitrary waveform: 1 Press Source; then, use the Entry knob to select the analog channel, math, or reference location whose waveform you wish to capture. 2 Press Store Source to Arb. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 259: To Output The Waveform Generator Sync Pulse

    (or the DC offset value). Cardiac DC, Noise, and Cardiac To specify the expected output load 1 If the Waveform Generator Menu is not currently displayed on the oscilloscope's softkeys, press the [Wave Gen] key. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 260: To Use Waveform Generator Logic Presets

    +2.5 V (TTL compatible) +5 V CMOS (5.0V) Not Available +5 V CMOS (3.3V) +2.5 V (CMOS +3.3 V compatible) CMOS (2.5V) +2.5 V +2.5 V -1.7 V -0.8 V (ECL compatible) -0.9 V Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 261: To Add Noise To The Waveform Generator Output

    3 In the Waveform Generator Settings Menu, press the Default Wave Gen softkey. The waveform generator factory default settings (1 kHz sine wave, 500 mVpp, 0 V offset, High- Z output load) are restored. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 262 Waveform Generator Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 263: Save/Recall (Setups, Screens, Data)

    PNG formats. Acquired waveform data can be saved to a USB storage device in comma- separated value (CSV), ASCII XY, binary (BIN), and Agilent logic analyzer binary (ALB) formats. There is also a command to securely erase all the oscilloscope's non- volatile internal memory.
  • Page 264 "To save reference waveform files to a USB storage device" on page 271. ALB data (*.alb) — This creates an Agilent proprietary format file that • can be imported by the Agilent Logic Analyzer application software, using the Agilent B4610A Data Import tool, for offline viewing and analysis.
  • Page 265: To Save Setup Files

    298. To save setup files Setup files can be saved to one of 10 internal (\Agilent Flash) locations or to an external USB storage device. 1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select Setup (*.scp).
  • Page 266: To Save Csv, Ascii Xy, Or Bin Data Files

    You can also save the oscilloscope's display image using a web browser. See "Get N O T E Image" on page 309 for details. To save CSV, ASCII XY, or BIN data files Data files can be saved to an external USB storage device. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 267: To Save Alb Data Files

    333 To save ALB data files Agilent logic analyzer binary (ALB) format data files can be viewed and analyzed offline on a PC using the Agilent Logic Analyzer application and the Agilent B4610A Data Import Tool. See the "Agilent Technologies...
  • Page 268 1 or If you connect your serial bus signals to the oscilloscope as shown in the following table, the bus/signal names in the Agilent Logic Analyzer application will be correct. Otherwise, you must re- map the signals in the Agilent Logic Analyzer application.
  • Page 269: Length Control

    ALB format files. It sets the number of data points that will be output to the file. Only displayed data points are saved. The maximum number of data points depends on these things: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 270 ASCII XY medium slowest See Also • "Binary Data (.bin) Format" on page 325 • "CSV and ASCII XY files" on page 332 • "Minimum and Maximum Values in CSV Files" on page 333 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 271: To Save Lister Data Files

    A message indicating whether the save was successful is displayed. To save masks Mask files can be saved to one of four internal (\Agilent Flash) locations or to an external USB storage device. 1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select Mask (*.msk).
  • Page 272: To Save Arbitrary Waveforms

    15, “Mask Testing,” starting on page 239 To save arbitrary waveforms Arbitrary waveform files can be saved to one of four internal (\Agilent Flash) locations or to an external USB storage device. 1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select Arbitrary Waveform data (*.csv).
  • Page 273: To Enter File Names

    It will truncate characters as necessary when the file name length is at maximum and more digits are required for the numeric portion of the file name. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 274: Recalling Setups, Masks, Or Data

    298. To recall setup files Setup files can be recalled from one of 10 internal (\Agilent Flash) locations or from an external USB storage device. 1 Press [Save/Recall] > Recall > Recall:; then, turn the Entry knob to select Setup (*.scp).
  • Page 275: To Recall Reference Waveform Files From A Usb Storage Device

    To recall arbitrary waveforms Arbitrary waveform files can be recalled from one of four internal (\ Agilent Flash) locations or from an external USB storage device. When recalling arbitrary waveforms (from an external USB storage device) that were not saved from the oscilloscope, be aware that: •...
  • Page 276: Recalling Default Setups

    Some user settings are not changed when recalling the default setup. Factory Default— recalls the oscilloscope's factory default settings. • You must confirm the recall because there are no user settings that are left unchanged. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 277: Performing A Secure Erase

    This performs a secure erase of all non- volatile memory in compliance with National Industrial Security Program Operation Manual (NISPOM) Chapter 8 requirements. You must confirm the secure erase, and the oscilloscope will reboot when finished. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 278 Save/Recall (Setups, Screens, Data) Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 279: Print (Screens)

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Print (Screens) To print the oscilloscope's display To set up network printer connections To specify the print options To specify the palette option You can print the complete display, including the status line and softkeys, to a USB printer or a network printer when the DSOXLAN LAN/VGA module is installed.
  • Page 280: To Set Up Network Printer Connections

    1 Press the [Print] key on the front panel. 2 In the Print Configuration Menu, press the Print to softkey; then, turn the Entry knob to select the network printer you want to configure (either #0 or #1). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 281: To Specify The Print Options

    In the Print Configuration Menu, press the Options softkey to change the following options: • Setup Information — Select this to print oscilloscope setup information on your printout, including vertical, horizontal, trigger, acquisition, math, and display settings. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 282: To Specify The Palette Option

    The oscilloscope's print driver cannot print color images to color laser printers, so the Color option is not available when connected to laser printers. • Grayscale — Select this to print the screen in shades of gray rather than in color. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 283: I/O Interface Settings

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide Utility Settings I/O Interface Settings Setting up the Oscilloscope's LAN Connection File Explorer Setting Oscilloscope Preferences Setting the Oscilloscope's Clock Setting the Rear Panel TRIG OUT Source Performing Service Tasks Configuring the [Quick Action] Key This chapter explains oscilloscope utility functions.
  • Page 284: Utility Settings

    When you connect the oscilloscope to a LAN it is a good practice to limit access to the N O T E oscilloscope by setting a password. By default, the oscilloscope is not password protected. "Setting a Password" on page 312 to set a password. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 285: To Establish A Lan Connection

    2 Press [Utility] > I/O. 3 Press the LAN Settings softkey. 4 Press the Config softkey; then, turn the Entry knob to select Automatic, and press the softkey again to disable it. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 286: Stand-Alone (Point-To-Point) Connection To A Pc

    LAN configuration settings 4 Connect your PC to the oscilloscope using a cross- over LAN cable such as Agilent part number 5061- 0701, available on the web at "www.parts.agilent.com". 5 Cycle power on the oscilloscope. Wait until the LAN connection is configured: Press [Utility] >...
  • Page 287: File Explorer

    The square receptacle on the rear panel labeled "DEVICE" is provided for controlling the oscilloscope over USB. See the Programmer's Guide for more information. The oscilloscope's internal file system, under "\Agilent Flash", consists of 10 locations for oscilloscope setup files, and four locations for mask files.
  • Page 288 When the USB mass storage device is connected to the oscilloscope's front or rear USB host port, a small four- color circle icon may be displayed briefly as the USB device is read. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 289: Setting Oscilloscope Preferences

    The ground level of the signal is identified by the position of the ground level ( ) icon at the far- left side of the display. The ground level will not move when you adjust the vertical sensitivity (volts/division) control. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 290: To Disable/Enable Transparent Backgrounds

    Use the Entry knob to choose a character. Then press the Enter softkey to advance to the next character and repeat the process. The resultant string is displayed in the "Text =" line above the softkeys. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 291: To Set Autoscale Preferences

    Press the Acq Mode softkey and turn the Entry knob to select whether • the acquisition mode should be preserved during autoscale: Normal — to make the oscilloscope switch to Normal acquisition • mode whenever the [AutoScale] key is pressed. This is the default mode. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 292: Setting The Oscilloscope's Clock

    TRIG OUT. The rising edge is delayed 30 ns from the oscilloscope's trigger point. The output level is 0- 5 V into an open circuit, and 0- 2.5 V into 50 . See Chapter 10, “Triggers,” starting on page 141. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 293: Performing Service Tasks

    "To clean the oscilloscope" on page 297 • "To check warranty and extended services status" on page 297 • "To contact Agilent" on page 297 • "To return the instrument" on page 298 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 294: To Perform User Calibration

    Performing User Cal will invalidate your Certificate of Calibration. If NIST (National Institute of Standards and Technology) traceability is required, perform the "Performance Verification" procedure in the Agilent InfiniiVision 2000/3000 X- Series Oscilloscopes Service Guide using traceable sources. To perform user calibration:...
  • Page 295 Then connect a BNC(f)- to- BNC(f) (barrel connector) to the tee as shown below. To Channel 1 To Channel 2 To Channel 3 Longer cable to TRIG OUT To Channel 4 Figure 49 User Calibration cable for 4-channel oscilloscope Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 296: To Perform Hardware Self Test

    Press [Help] > About Oscilloscope to display information about your oscilloscope: • Model number. • Serial number. • Bandwidth. • Module installed. • Software version. • Installed licenses. See also "Loading Licenses and Displaying License Information" on page 323. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 297: To Display The User Calibration Status

    If the system cannot find your product's warranty status, select Contact Us and speak with an Agilent Technologies representative. To contact Agilent Information on contacting Agilent Technologies can be found at: "www.agilent.com/find/contactus" Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 298: Measurements," Starting On

    Utility Settings To return the instrument Before shipping the oscilloscope to Agilent Technologies, contact your nearest Agilent Technologies sales or service office for additional details. Information on contacting Agilent Technologies can be found at: "www.agilent.com/find/contactus" 1 Write the following information on a tag and attach it to the oscilloscope.
  • Page 299: Print (Screens)," Starting On

    178. Quick Clear Display — clears the display, see • "To clear the display" page 132. Once the [Quick Action] key is configured, you simply press it to perform the selected action. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 300 Utility Settings Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 301: Web Interface

    Identification Function Instrument Utilities Setting a Password When the Agilent InfiniiVision X- Series oscilloscopes have the DSOXLAN LAN/VGA option module installed, you can access the oscilloscope's built- in web server using a Java- enabled web browser. The oscilloscope's web interface lets you: •...
  • Page 302: Accessing The Web Interface

    It is possible to use a point- to- point connection, but using a normal LAN connection is the preferred method. 2 Type the oscilloscope's hostname or IP address in the web browser. The oscilloscope's web interface Welcome Page is displayed. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 303: Browser Web Control

    If Java is not installed on your PC, you will be prompted to install the Sun Microsystems N O T E Java Plug-in. This plug-in must be installed on the controlling PC for the web interface's Remote Front Panel or Remote Programming operations. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 304: Real Scope Remote Front Panel

    The SCPI Command window is useful for testing commands or entering a few commands interactively. When creating automated programs for controlling the oscilloscope, you will typically use the Agilent IO Libraries from within a programming environment like Microsoft Visual Studio (see "Remote Programming with Agilent IO Libraries"...
  • Page 305: Simple Remote Front Panel

    To Resolution display the remote front panel without scroll bars, use a monitor resolution greater than 800 x 600 on your computer's display. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 306: Remote Programming Via The Web Interface

    "Accessing the Web Interface" on page 302). 2 When the oscilloscope's web interface is displayed, select Browser Web Control, then select Remote Programming. The SCPI Commands applet appears within the browser web page. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 307: Remote Programming With Agilent Io Libraries

    While the SCPI Commands applet window lets you enter and remote programming commands, remote programming for automated test and data acquisition is typically done using the Agilent IO Libraries, which are separate from the instrument's web interface. The Agilent IO Libraries let a controller PC communicate with Agilent...
  • Page 308 The current acquisition is saved. d In the File Download dialog, click Save. e In the Save As dialog, navigate to the folder where you want to save the file; then, click Save. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 309: Recalling Files Via The Web Interface

    When recalling reference waveform data files, select the To Reference Waveform location. d Click Recall. Get Image To save (or print) the oscilloscope's display from the web interface: 1 Access the oscilloscope's web interface (see "Accessing the Web Interface" on page 302). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 310: Identification Function

    2 When the oscilloscope's web interface Welcome Page is displayed, select the Identification on radio button. An "Identify" message is displayed on the oscilloscope; you can either select Identification off or press the OK softkey on the oscilloscope to continue. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 311: Instrument Utilities

    The Instrument Utilities page of the web interface lets you: • View installed options. • View firmware versions. • Install firmware upgrade files. • View calibration status. You can select these cababilities via a drop- down menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 312: Setting A Password

    To set a password: 1 Access the oscilloscope's web interface (see "Accessing the Web Interface" on page 302). 2 When the oscilloscope's web interface is displayed, select the Configure Network tab from the instrument's Welcome page. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 313 Web Interface 3 Click the Modify Configuration button. Modify Configuration Configure Network 4 Enter your desired password, and click Apply Changes. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 314 • Using the keys on the front panel of the oscilloscope, press [Utility] > I/O > LAN Reset. • Using the web browser select the Configure Network tab, select Modify Configuration, erase the Password, and select Apply Changes. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 315: Reference

    Or, go to the Agilent home page at "www.agilent.com" and search for "3000 X- Series oscilloscopes data sheet". To order a data sheet by phone, please contact your local Agilent office. The complete list is available at: "www.agilent.com/find/contactus". Measurement Category •...
  • Page 316: Oscilloscope Measurement Category

    Measurement category IV is for measurements performed at the source of the low- voltage installation. Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 317: Transient Withstand Capability

    This product is intended to be powered by MAINS that comply to Overvoltage Category Category II, which is typical of cord-and-plug connected equipment. Pollution Degree The InfiniiVision 2000/3000 X-Series oscilloscopes may be operated in environments of Pollution Degree 2 (or Pollution Degree 1). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 318: Probes And Accessories

    • "Accessories Available" on page 322 AutoProbe Most Agilent single- ended active, differential, and current probes are Interface compatible with the AutoProbe interface. Active probes that do not have their own external power supply require substantial power from the AutoProbe interface.
  • Page 319: Passive Probes

    These probes have a pin on their connector that connects to the ring around the oscilloscope's BNC connector. Therefore, the oscilloscope will automatically set the attenuation factor for recognized Agilent passive probes. Passive probes that do not have a pin that connects to the ring around the BNC connector will not be recognized by the oscilloscope, and you must set the probe attenuation factor manually.
  • Page 320: Differential Probes

    Active probe, 800 MHz (requires 1142A power supply) 1145A Active probe, 750 MHz 2-ch (requires 1142A power supply) N2772A Active differential probe, 20 MHz, 1.2 kVDC + peak AC max (requires N2773A power supply) Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 321: Current Probes

    Current probe, 50 MHz, 30 A, AC/DC (use with N2779A power supply) N2783A Current probe, 100 MHz, 30 A, AC/DC (use with N2779A power supply) N2893A Current probe, 100 MHz, 15 A, AC/DC with AutoProbe interface "AutoProbe Interface" on page 318. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 322: Accessories Available

    320), and current probes ("Current Probes" on page 321), the following accessories are available for the InfiniiVision 3000 X- Series oscilloscopes. Table 10 Accessories Available for InfiniiVision 3000 X-Series Oscilloscopes Model/Part # Description DSOXLAN LAN/VGA connection module DSOXGPIB...
  • Page 323: Loading Licenses And Displaying License Information

    RS232 (Recommended Standard 232). Educator's Kit Order DSOXEDK. Provides training signals on the oscilloscope's Demo terminals and a lab guide/tutorial for education environments. EMBD Embedded Serial Triggering and Analysis (I2C, SPI). Order DSOX3EMBD. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 324 "inactivity". Extended Video Triggering and Analysis. Order DSOX3VID. WAVEGEN Waveform Generator. Order DSOX3WAVEGEN. Other Options Available Table 12 Calibration Option Option Order ANSI Z540 Compliant Calibration Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 325: Software And Firmware Updates

    Agilent Technologies representative or see "www.agilent.com/find/3000X- Series". Software and Firmware Updates From time to time Agilent Technologies releases software and firmware updates for its products. To search for firmware updates for your oscilloscope, direct your web browser to "www.agilent.com/find/3000X- Series- sw".
  • Page 326: Binary Data In Matlab

    MathWorks MATLAB®. You can download the appropriate MATLAB functions from the Agilent Technologies web site at "www.agilent.com/find/3000X- Series- sw". Agilent provides the .m files, which need to be copied into the work directory for MATLAB. The default work directory is C:\MATLAB7\work. Binary Header Format File Header There is only one file header in a binary file.
  • Page 327 Reference Cookie Two byte characters, AG, that indicate the file is in the Agilent Binary Data file format. Version Two bytes that represent the file version. File Size A 32-bit integer that is the number of bytes that are in the file.
  • Page 328 A waveform may have more than one data set. Each waveform data set Header will have a waveform data header. The waveform data header consists of information about the waveform data set. This header is stored immediately before the data set. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 329: Example Program For Reading Binary Data

    "Example Program for Reading Binary Data". Examples of Binary Files Single The following picture shows a binary file of a single acquisition with Acquisition multiple analog channels. Multiple Analog Channels Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 330 Bytes per Point = 4 12 bytes Voltage Data N buffer size Single The following picture shows a binary file of a single acquisition with all Acquisition All pods for the logic channels saved. Pods Logic Channels Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 331 Bytes per Point = 1 12 bytes Pod 2 Timing Data buffer size Segmented The following picture shows a binary file of a segmented memory Memory acquisition on one analog channel. Acquisition on One Analog Channel Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 332: Csv And Ascii Xy Files

    12 bytes Voltage Data N buffer size CSV and ASCII XY files • "CSV and ASCII XY file structure" on page 333 • "Minimum and Maximum Values in CSV Files" on page 333 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 333: Csv And Ascii Xy File Structure

    CSV record is configured to provide a best- estimate of all the samples that each point in the record represents. Therefore, the min and max values may not appear in the CSV file. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 334: Acknowledgements

    GNU General Public License for more details. The license is located on the Agilent InfiniiVision Oscilloscopes Documentation CD- ROM. RealVNC source code can be obtained from RealVNC or by contacting Agilent. Agilent will charge for the cost of physically performing the source distribution. HDF5 Reference Waveform files use HDF5.
  • Page 335: Can/Lin Triggering And Serial Decode

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide CAN/LIN Triggering and Serial Decode Setup for CAN Signals CAN Triggering CAN Serial Decode Setup for LIN Signals LIN Triggering LIN Serial Decode CAN/LIN triggering and serial decode requires Option AUTO or the DSOX3AUTO upgrade.
  • Page 336 10 Press the Signal softkey and select the type and polarity of the CAN signal. This also automatically sets the channel label for the source channel. CAN_H — The actual CAN_H differential bus. • Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 337: Can Triggering

    1 Press [Trigger]. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select the serial slot (Serial 1 or Serial 2) on which the CAN signal is being decoded. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 338 If the setup does not produce a stable trigger, the CAN signal may be slow enough that the N O T E oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 339: Can Serial Decode

    See Also • "Interpreting CAN Decode" on page 340 • "CAN Totalizer" on page 341 • "Interpreting CAN Lister Data" on page 342 • "Searching for CAN Data in the Lister" on page 343 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 340: Interpreting Can Decode

    • Pink vertical bars indicate you need to expand the horizontal scale (and run again) to see decode. • Red dots in the decode line indicate that there is data that is not being displayed. Scroll or expand the horizontal scale to view the information. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 341: Can Totalizer

    • The UTIL (bus load) indicator measures the percentage of time the bus is active. The calculation is done on 330 ms periods, approximately every 400 ms. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 342: Interpreting Can Lister Data

    (Fo), or Frame (Fr). Different kinds of errors can be combined like "Fo,Fr" in the above example. Aliased data is highlighted in pink. When this happens, decrease the horizontal time/div setting and run again. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 343: Searching For Can Data In The Lister

    LIN signal, specifying the signal source, threshold voltage level, baud rate, sample point, and other LIN signal parameters. To set up the oscilloscope to capture LIN signals: 1 Press [Label] to turn on labels. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 344 Back/Up key to return to the LIN Signals Menu. 11 Press the Sample Point softkey; then, turn the Entry knob to select the sample point at which the oscilloscope will sample the bit value. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 345: Lin Triggering

    1 Press [Trigger]. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select the serial slot (Serial 1 or Serial 2) on which the CAN signal is being decoded. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 346 For details about using the LIN Bits Menu softkeys, press and hold the softkey in question N O T E to display the built-in help. For LIN decode information see "LIN Serial Decode" on page 347. N O T E Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 347: Lin Serial Decode

    You can use the horizontal Zoom window for easier navigation of the decoded data. See Also • "Interpreting LIN Decode" on page 348 • "Interpreting LIN Lister Data" on page 349 • "Searching for LIN Data in the Lister" on page 350 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 348: Interpreting Lin Decode

    Scroll or expand the horizontal scale to view the information. • Unknown bus values (undefined or error conditions) are drawn in red. • If there is an error in the synch field, SYNC will appear in red. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 349: Interpreting Lin Lister Data

    • Checksum — (LIN 1.3 only). • Data and Checksum — (LIN 2.0 only). • Errors — highlighted in red. Aliased data is highlighted in pink. When this happens, decrease the horizontal time/div setting and run again. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 350: Searching For Lin Data In The Lister

    Errors — Finds all errors. • For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 351: Flexray Triggering And Serial Decode

    Setup for FlexRay Signals FlexRay signal setup consists of first connecting the oscilloscope to a differential FlexRay signal using a differential active probe (the Agilent N2792A is recommended), specifying the signal source, threshold voltage trigger level, baud rate, and bus type.
  • Page 352: Flexray Triggering

    (see page 353), errors (see page 354), or events (see page 354). To display FlexRay serial decode, see "FlexRay Serial Decode" on page 355. N O T E Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 353: Triggering On Flexray Frames

    0 through the Cyc Ct Rep factor minus 1. For example, with a base factor of 1 and a repetition factor of 16, the oscilloscope triggers on cycles 1, 17, 33, 49, and 65. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 354: Triggering On Flexray Errors

    Normal. This prevents the oscilloscope from Auto triggering while waiting for an error to occur. You may need to adjust trigger holdoff to see a particular error when multiple errors exist. Triggering on FlexRay Events 1 Press [Trigger]. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 355: Flexray Serial Decode

    To set up FlexRay serial decode: 1 Press [Serial] to display the Serial Decode Menu. 2 If the decode line does not appear on the display, press the [Serial] key to turn it on. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 356: Interpreting Flexray Decode

    • Payload- length (decimal number of words in green). • Header CRC (hex digits in blue plus red HCRC error message if invalid). • Cycle number (decimal digits in yellow). • Data bytes (HEX digits in white). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 357: Flexray Totalizer

    The totalizer runs, counting frames and calculating percentages, even when the oscilloscope is stopped (not acquiring data). When an overflow condition occurs, the counter displays OVERFLOW. The counters can be reset to zero by pressing the Reset FlexRay Counters softkey. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 358: Interpreting Flexray Lister Data

    The oscilloscope's search capability lets you search for (and mark) certain types of FlexRay data in the Lister. You can use the [Navigate] key and controls to navigate through the marked rows. 1 With FlexRay selected as the serial decode mode, press [Search]. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 359 Errors — finds all errors. • For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 360 FlexRay Triggering and Serial Decode Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 361: I2C/Spi Triggering And Serial Decode

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide I2C/SPI Triggering and Serial Decode Setup for I2C Signals I2C Triggering I2C Serial Decode Setup for SPI Signals SPI Triggering SPI Serial Decode I2C/SPI triggering and serial decode requires Option EMBD or the DSOX3EMBD upgrade.
  • Page 362: I2C Triggering

    EEPROM data read, or on a read/write frame with a specific device address and data value. 1 Press [Trigger]; then, select the I2C trigger type. 2 Press [Trigger]. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 363 Data is softkey. When this event occurs, the oscilloscope will trigger on the clock edge for the Ack bit after the data byte. This data byte does not need to occur directly after the control byte. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 364 26th clock edge if all bits in the pattern match. The frame is in the format: Frame (Start: Address byte 1: Write: Address byte 2: Ack: Data) For triggering purposes, a restart is treated as a start condition. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 365 8- bit data pattern on which to trigger. You can select a data value in the range of 0x00 to 0xFF (hexadecimal). The oscilloscope will trigger after the start, address, read/write, acknowledge, and data events occur. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 366: I2C Serial Decode

    3 If the decode line does not appear on the display, press the [Serial] key to turn it on. 4 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 367: Interpreting I2C Decode

    • Address values appear at the start of a frame. • Write addresses appear in light- blue along with the "W" character. • Read addresses appear in yellow along with the "R" character. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 368: Interpreting I2C Lister Data

    In addition to the standard Time column, the I2C Lister contains these columns: • Restart — indicated with an "X". • Address — colored blue for writes, yellow for reads. • Data — data bytes. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 369: Searching For I2C Data In The Lister

    26th clock edge if all bits in the pattern match. For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 370: Setup For Spi Signals

    Press the slope softkey ( ) to select rising edge or falling edge for the selected Clock source. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 371 Press the Threshold softkey; then, turn the Entry knob to select the MISO signal threshold voltage level. The threshold voltage level is used in decoding, and it will become the trigger level when the trigger type is set to the selected serial decode slot. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 372 The framing signal must be valid for the entire data pattern. If the framing signal is set to Timeout, the oscilloscope generates • it's own internal framing signal after it sees inactivity on the serial clock line. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 373: Spi Triggering

    1 Press [Trigger]. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select the serial slot (Serial 1 or Serial 2) on which the SPI signals are being decoded. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 374 As you rotate the Entry knob, the bit is highlighted in the Data string shown in the waveform area. b Press the 0 1 X softkey to set the bit selected in the Bit softkey to 0 (low), 1 (high), or X (don't care). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 375: Spi Serial Decode

    Signals" on page 370. For SPI triggering setup see "SPI Triggering" on page 373. N O T E To set up SPI serial decode: 1 Press [Serial] to display the Serial Decode Menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 376 You can use the horizontal Zoom window for easier navigation of the acquired data. See Also • "Interpreting SPI Decode" on page 377 • "Interpreting SPI Lister Data" on page 378 • "Searching for SPI Data in the Lister" on page 378 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 377: Interpreting Spi Decode

    Scroll or expand the horizontal scale to view the information. • Aliased bus values (undersampled or indeterminate) are drawn in pink. • Unknown bus values (undefined or error conditions) are drawn in red. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 378: Interpreting Spi Lister Data

    (Serial 1 or Serial 2) on which the SPI signals are being decoded. 3 Press Search; then, select from these options: • Master-Out, Slave-In (MOSI) Data — for searching MOSI data. • Master-In, Slave-Out (MISO) Data — for searching MISO data. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 379 For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 380 I2C/SPI Triggering and Serial Decode Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 381: I2S Triggering And Serial Decode

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide I2S Triggering and Serial Decode Setup for I2S Signals I2S Triggering I2S Serial Decode I2S triggering and serial decode requires Option AUDIO or the DSOX3AUDIO upgrade. Only one I2S serial bus can be decoded at a time.
  • Page 382 (from 4 to 32 bits). 11 Press the Alignment softkey; then, turn the Entry knob to select the desired alignment of the data signal (SDATA). The on- screen diagram changes with your selection. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 383 WS Low indicates Left or Right channel data. The on- screen diagram changes with your selection. WS Low = Left Channel — Left- channel data corresponds to WS=low; right- channel data corresponds to WS=high. WS Low=Left is the oscilloscope's default WS setting. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 384: I2S Triggering

    1 Press [Trigger]. 2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob to select the serial slot (Serial 1 or Serial 2) on which the I2S signals are being decoded. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 385 2 points shown since there are two instances in which the data increases from a value below (or at) the "Armed" value to a value greater than (or equal to) the specified "Trigger" value. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 386 If the setup does not produce a stable trigger, the I2S signal may be slow enough that the N O T E oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode softkey to set the trigger mode from Auto to Normal. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 387: I2S Serial Decode

    Auto to Normal. You can use the horizontal Zoom window for easier navigation of the acquired data. See Also • "Interpreting I2S Decode" on page 388 • "Interpreting I2S Lister Data" on page 389 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 388: Interpreting I2S Decode

    Scroll or expand the horizontal scale to view the data. • Aliased bus values (undersampled or indeterminate) are drawn in pink. • Unknown bus values (undefined or error conditions) are drawn in red. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 389: Interpreting I2S Lister Data

    The oscilloscope's search capability lets you search for (and mark) certain types of I2S data in the Lister. You can use the [Navigate] key and controls to navigate through the marked rows. 1 With I2S selected as the serial decode mode, press [Search]. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 390 Errors — finds all errors. • For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 391: Mil-Std-1553/Arinc 429 Triggering And Serial Decode

    Setup for MIL-STD-1553 Signals MIL- STD- 1553 signal setup consists of first connecting the oscilloscope to a serial MIL- STD- 1553 signal using a differential active probe (the Agilent N2791A is recommended), specifying the signal source, and specifying the high and low trigger threshold voltage levels.
  • Page 392 Press the Low Threshold softkey; then, turn the Entry knob to set the low trigger threshold voltage level. The threshold voltage levels are used in decoding and will become the trigger levels when the trigger type is set to the selected serial decode slot. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 393: Mil-Std-1553 Triggering

    When this option is selected, these softkeys become available: The RTA softkey lets you select the hex Remote Terminal Address • value. The Bit Time softkey lets you select the bit time position. • Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 394: Mil-Std-1553 Serial Decode

    3 If the decode line does not appear on the display, press the [Serial] key to turn it on. 4 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode data. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 395: Interpreting Mil-Std-1553 Decode

    • Data word decoded data is colored white, preceded by the text "D:". • Command/Status or Data words with a Parity error have the decode text displayed in red instead of green or white. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 396: Interpreting Mil-Std-1553 Lister Data

    • Data — the 11 bits after the RTA for Command/Status words, or the 16 bits of a Data word. • Errors — "Sync", "Parity", or "Manchester" errors as appropriate. The background color is red to indicate an error. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 397: Searching For Mil-Std-1553 Data In The Lister

    Parity Error — finds (odd) parity bits that are incorrect for the data in • the word. Sync Error — finds invalid Sync pulses. • Manchester Error — finds Manchester encoding errors. • For more information on searching data, see "Searching Lister Data" page 126. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 398: Setup For Arinc 429 Signals

    Setup for ARINC 429 Signals Setup consists of first connecting the oscilloscope to a ARINC 429 signal using a differential active probe (the Agilent N2791A is recommended), then using the Signals Menu to specify the signal source, the high and low trigger threshold voltage levels, the signal speed, and the signal type.
  • Page 399: Arinc 429 Triggering

    (Serial 1 or Serial 2) on which the ARINC 429 signal is being decoded. 3 Press the Trigger: softkey; then, turn the Entry knob to select the trigger condition: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 400 Use the Set all Bits softkey to set all bit values to 0, 1, or X. 6 If you select the Label Range condition, use the Lable Min and Label Max softkeys to specify the ends of the range. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 401: Arinc 429 Serial Decode

    3 Press the Word Format softkey and specify the word decode format: Label/SDI/Data/SSM: • • Label - 8 bits. • SDI - 2 bits. • Data - 19 bits. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 402 "Interpreting ARINC 429 Decode" on page 403 • "ARINC 429 Totalizer" on page 404 • "Interpreting ARINC 429 Lister Data" on page 405 • "Searching for ARINC 429 Data in the Lister" on page 405 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 403: Interpreting Arinc 429 Decode

    Data (white, red if parity error) (21 bits) – displayed in the selected Base. • SSM (green) (2 bits) – displayed in binary. • When the decode format is Label/Data: • Label (yellow) (8 bits) – displayed in octal. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 404: Arinc 429 Totalizer

    Pressing the [Run/Stop] key does not affect the totalizer. When an overflow condition occurs, the counter displays OVERFLOW. The counters can be reset to zero by pressing the Reset ARINC 429 Counters softkey (in the decode Settings menu). Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 405: Interpreting Arinc 429 Lister Data

    ARINC 429 data in the Lister. You can use the [Navigate] key and controls to navigate through the marked rows. 1 With ARINC 429 selected as the serial decode mode, press [Search]. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 406 All Errors — finds any of the above errors. • For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 407: Uart/Rs232 Triggering And Serial Decode

    Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide UART/RS232 Triggering and Serial Decode Setup for UART/RS232 Signals UART/RS232 Triggering UART/RS232 Serial Decode UART/RS232 triggering and serial decode requires Option COMP or the DSOX3COMP upgrade. Setup for UART/RS232 Signals To set up the oscilloscope to capture UART/RS232 signals: 1 Press [Label] to turn on labels.
  • Page 408 Bit Order — Select whether the most significant bit (MSB) or the least significant bit (LSB) is presented after the start bit in the signal from your device under test. For RS232 select LSB. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 409: Uart/Rs232 Triggering

    Note that the setting of this softkey does not affect the selected base of the decode display. 5 Press the Trigger softkey and set up the desired trigger condition: Rx Start Bit — The oscilloscope triggers when a start bit occurs on Rx. • Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 410 9 If Burst is selected, an idle time (1 µs to 10 s) can be specified so that the oscilloscope will look for a trigger condition only after the idle time has past. Press the Idle softkey and rotate the Entry knob to set an idle time. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 411: Uart/Rs232 Serial Decode

    Valid ASCII characters are between 0x00 and 0x7F. To display in ASCII you must select at least 7 bits in the Bus Configuration. If ASCII is selected and the data exceeds 0x7F, the data is displayed in hex. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 412 See Also • "Interpreting UART/RS232 Decode" on page 413 • "UART/RS232 Totalizer" on page 414 • "Interpreting UART/RS232 Lister Data" on page 415 • "Searching for UART/RS232 Data in the Lister" on page 415 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 413: Interpreting Uart/Rs232 Decode

    Expand the horizontal scale to allow the data to display. • An unknown (undefined) bus is shown in red. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 414: Uart/Rs232 Totalizer

    When an overflow condition occurs, the counter displays OVERFLOW. The counters can be reset to zero by pressing the Reset UART Counters softkey in the UART/RS232 Settings Menu. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 415: Interpreting Uart/Rs232 Lister Data

    2 In the Search Menu, press the Search softkey; then, turn the Entry knob to select the serial slot (Serial 1 or Serial 2) on which the UART/RS232 signals are being decoded. 3 In the Search Menu, press Search; then, select from these options: Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 416 Rx or Tx Any Error — Finds any error. • For more information on searching data, see "Searching Lister Data" page 126. For more information on using the [Navigate] key and controls, see "Navigating the Time Base" on page 58. Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 417 Auto? trigger indicator, trigger, AutoIP, 285, Addresses softkey, automatic measurements, 213, calibrate probe, ADVMATH license, automatic setup, calibration, AERO license, AutoProbe interface, 39, calibration protect button, 42, Agilent IO Libraries Suite, AutoScale preferences, calibration status, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 418 63, coupling, trigger, differentiate math function, expand about center, crosstalk problems, Digit softkey, expand about ground, CSV file format, digital bus mode, exponential fall waveform generator digital channel controls, output, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 419 Ground terminal, Instrument Utilities web page, flash drive, Integrate math function, Flat top FFT window, intensity control, FLEX license, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 420 323, measurement category, definitions, network, connecting to, LIN serial decode, measurement definitions, new label, LIN trigger, measurement record, noise rejection, line voltage, measurement statistics, noise waveform generator output, Lister, measurement thresholds, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 421 Quick Recall, password (network), setting, probe head, Rectangular FFT window, pattern trigger, probe units, Ref key, 37, pattern, SPI trigger, probe, AutoProbe interface, reference point, waveform, PC connection, probe, calibrate, reference waveforms, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 422 SCL, I2C trigger, size, time reference indicator, SCLK, I2S trigger, skew, analog channel, time, re-arm, SCPI Commands window, slope trigger, timebase, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...
  • Page 423 RS232, usb2, waveform type, waveform generator, trigger type, runt, user cal, waveform, cursor tracking, trigger type, setup and hold, user calibration, waveform, intensity, trigger type, slope, user interface language, waveform, printing, Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide...