25
:POWer Commands

:POWer:SWITch:IREFerence

Command Syntax
:POWer:SWITch:IREFerence <percent>
<percent> ::= percent in NR1 format
The :POWer:SWITch:IREFerence command to specify the current switching
level for the start of switching edges. The value is in percentage of the
maximum switch current.
You can adjust this value to ignore noise floors or null offset that is
difficult to eliminate in current probes.
This value specifies the threshold that is used to determine the switching
edges.
Query Syntax
:POWer:SWITch:IREFerence?
The :POWer:SWITch:IREFerence query returns the current switching level
percent value.
Return Format
<percent><NL>
<percent> ::= percent in NR1 format
See Also
604
(see
page
1208)
":POWer:SWITch:APPLy"
":POWer:SWITch:CONDuction"
":POWer:SWITch:RDS"
":POWer:SWITch:VCE"
":POWer:SWITch:VREFerence"