Measure:vpp - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

21
:MEASure Commands

:MEASure:VPP

Command Syntax
:MEASure:VPP [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:VPP command installs a screen measurement and starts a
vertical peak- to- peak measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VPP? [<source>]
The :MEASure:VPP? query measures the maximum and minimum vertical
value for the selected source, then calculates the vertical peak- to- peak
value and returns that value. The peak- to- peak value (Vpp) is calculated
with the following formula:
Vmax and Vmin are the vertical maximum and minimum values present on
the selected source.
Return Format
<value><NL>
<value> ::= vertical peak to peak value in NR3 format
See Also
476
(see
page
1208)
Vpp = Vmax - Vmin
"Introduction to :MEASure Commands"
":MEASure:SOURce"
":MEASure:VMAX"
on page 474
":MEASure:VMIN"
on page 475
":MEASure:VAMPlitude"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 420
on page 458
on page 471

Advertisement

Table of Contents
loading

Table of Contents