Trigger:runt:polarity - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

:TRIGger:RUNT:POLarity

Command Syntax
:TRIGger:RUNT:POLarity <polarity>
<polarity> ::= {POSitive | NEGative | EITHer}
The :TRIGger:RUNT:POLarity command sets the polarity for the runt
trigger:
• POSitive — positive runt pulses.
• NEGative — negative runt pulses.
• EITHer — either positive or negative runt pulses.
Query Syntax
:TRIGger:RUNT:POLarity?
The :TRIGger:RUNT:POLarity? query returns the runt trigger polarity.
Return Format
<polarity><NL>
<polarity> ::= {POS | NEG | EITH}
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 944
":TRIGger:LEVel:HIGH"
":TRIGger:LEVel:LOW"
":TRIGger:RUNT:SOURce"
on page 935
on page 942
on page 943
on page 989