22
:MEASure Power Commands

:MEASure:APParent

Command Syntax
:MEASure:APParent [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:APParent command installs an apparent power
measurement on screen.
The <source1> parameter is the channel probing voltage and the <source2>
parameter is the channel probing current. These sources can also be
specified by the :MEASure:SOURce command.
Apparent power is a measure of power quality. It is the portion of AC line
power flow due to stored energy which returns to the source in each
cycle.
IRMS * VRMS
Query Syntax
:MEASure:APParent? [<source1>][,<source2>]
The :MEASure:APParent query returns the measured apparent power.
Return Format
<return_value><NL>
<return_value> ::= the apparent power value in NR3 format
See Also
490
(see
page
1208)
":MEASure:SOURce"
":POWer:QUALity:APPLy"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 458
on page 579