:MTESt Commands

:MTESt:SCALe:Y2

Command Syntax
:MTESt:SCALe:Y2 <y2_value>
<y2_value> ::= Y2 value in NR3 format
The :MTESt:SCALe:Y2 command defines the Y2 marker in the coordinate
system for mask testing. All Y values of vertices in the coordinate system
are defined with respect to the boundaries defined by SCALe:Y1 and
SCALe:Y2 according to the following equation:
Y = (Y * (Y2 - Y1)) + Y1
Thus, if you set Y1 to 100 mV, and Y2 to 1 V, a Y value of 0.100 in a
vertex is at 190 mV.
The Y2 value is a voltage value specifying the location of the Y2 marker.
Query Syntax
:MTESt:SCALe:Y2?
The :MTESt:SCALe:Y2? query returns the current setting of the Y2 marker.
Return Format
<y2_value><NL>
<y2_value> ::= Y2 value in NR3 format
See Also
538
(see
page
1208)
"Introduction to :MTESt Commands"
":MTESt:SCALe:BIND"
":MTESt:SCALe:X1"
on page 535
":MTESt:SCALe:XDELta"
":MTESt:SCALe:Y1"
on page 537
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 509
on page 534
on page 536