:HWEenable (Hardware Event Enable Register)
Command Syntax
:HWEenable <mask>
<mask> ::= 16-bit integer
The :HWEenable command sets a mask in the Hardware Event Enable
register. Set any of the following bits to "1" to enable bit 12 in the
Operation Status Condition Register and potentially cause an SRQ (Service
Request interrupt to be generated.
Table 72 Hardware Event Enable Register (HWEenable)
Bit
15-13
12
11-0
Query Syntax
:HWEenable?
The :HWEenable? query returns the current value contained in the
Hardware Event Enable register as an integer number.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
Name
Description
---
---
PLL
PLL Locked
Locked
---
---
"Introduction to Root (:) Commands"
":AER (Arm Event Register)"
":CHANnel<n>:PROTection"
":OPERegister[:EVENt] (Operation Status Event Register)"
":OVLenable (Overload Event Enable Register)"
When Set (1 = High = True), Enables:
(Not used.)
This bit is for internal use and is not intended
for general use.
(Not used.)
on page 190
on page 192
on page 278
Root (:) Commands
on page 213
on page 215
6