Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual page 243

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

Control all oscilloscope functions associated with buses made up of digital
channels. See
Table 83 :BUS<n> Commands Summary
Command
:BUS<n>:BIT<m> {{0 |
OFF} | {1 | ON}} (see
page
245)
:BUS<n>:BITS
<channel_list>, {{0 |
OFF} | {1 | ON}} (see
page
246)
:BUS<n>:CLEar (see
page
248)
:BUS<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
page
249)
Agilent InfiniiVision 4000 X-Series Oscilloscopes
Programmer's Guide
8
:BUS<n> Commands
"Introduction to :BUS<n> Commands"
Query
:BUS<n>:BIT<m>? (see
page
245)
:BUS<n>:BITS? (see
page
246)
n/a
:BUS<n>:DISPlay? (see
page
249)
s1
on page 244.
Options and Query Returns
{0 | 1}
<n> ::= 1 or 2; an integer in NR1
format
<m> ::= 0-15; an integer in NR1
format
<channel_list>, {0 | 1}
<channel_list> ::= (@<m>,<m>:<m>
...) where "," is separator and
":" is range
<n> ::= 1 or 2; an integer in NR1
format
<m> ::= 0-15; an integer in NR1
format
<n> ::= 1 or 2; an integer in NR1
format
{0 | 1}
<n> ::= 1 or 2; an integer in NR1
format
243

Advertisement

Table of Contents
loading

Table of Contents