Measure:statistics:increment - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

21
:MEASure Commands

:MEASure:STATistics:INCRement

Command Syntax
:MEASure:STATistics:INCRement
This command updates the statistics once (incrementing the count by one)
using the current measurement values. It corresponds to the front panel
Increment Statistics softkey in the Measurement Statistics Menu. This
command lets you, for example, gather statistics over multiple pulses
captured in a single acquisition. To do this, change the horizontal position
and enter the command for each new pulse that is measured.
This command is only allowed when the oscilloscope is stopped and quick
measurements are on.
The command is allowed in segmented acquisition mode even though the
corresponding front panel softkey is not available.
See Also
462
(see
page
1208)
"Introduction to :MEASure Commands"
":MEASure:STATistics"
":MEASure:STATistics:DISPlay"
":MEASure:STATistics:RESet"
":MEASure:RESults"
on page 452
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 420
on page 460
on page 461
on page 464