21
:MEASure Commands

:MEASure:RESults

Query Syntax
:MEASure:RESults?
The :MEASure:RESults? query returns the results of the continuously
displayed measurements. The response to the MEASure:RESults? query is a
list of comma- separated values.
If more than one measurement is running continuously, the
:MEASure:RESults return values are duplicated for each continuous
measurement from the first to last (top to bottom) result displayed. Each
result returned is separated from the previous result by a comma. There is
a maximum of 10 continuous measurements that can be continuously
displayed at a time.
When no quick measurements are installed, the :MEASure:RESults? query
returns nothing (empty string). When the count for any of the
measurements is 0, the value of infinity (9.9E+37) is returned for the min,
max, mean, and standard deviation.
Return Format
<result_list><NL>
<result_list> ::= comma-separated list of measurement results
The following shows the order of values received for a single measurement
if :MEASure:STATistics is set to ON.
Measurem
ent label
Measurement label, current, min, max, mean, std dev, and count are only
returned if :MEASure:STATistics is ON.
If :MEASure:STATistics is set to CURRent, MIN, MAX, MEAN, STDDev, or
COUNt only that particular statistic value is returned for each
measurement that is on.
See Also
Example Code
' This program shows the InfiniiVision oscilloscopes' measurement
' statistics commands.
' -------------------------------------------------------------------
Option Explicit
Public myMgr As VisaComLib.ResourceManager
Public myScope As VisaComLib.FormattedIO488
Public varQueryResult As Variant
Public strQueryResult As String
452
(see
page
1208)
current
min
"Introduction to :MEASure Commands"
":MEASure:STATistics"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
max
mean
on page 420
on page 460
std dev
count