Wgen<W>:Voltage - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

34
:WGEN<w> Commands
:WGEN<w>:VOLTage
Command Syntax
:WGEN<w>:VOLTage <amplitude>
<w> ::= 1 or 2 in NR1 format
<amplitude> ::= amplitude in volts in NR3 format
For all waveforms except DC, the :WGEN<w>:VOLTage command specifies
the waveform's amplitude. Use the :WGEN<w>:VOLTage:OFFSet command
to specify the offset voltage or DC level.
You can also specify the amplitude and offset indirectly using the
:WGEN<w>:VOLTage:HIGH and :WGEN<w>:VOLTage:LOW commands. For
example, an amplitude of 5 V and an offset of 1 V is the same as a
high- level voltage of 4 V and a low- level voltage of - 1 V.
Query Syntax
:WGEN<w>:VOLTage?
The :WGEN<w>:VOLTage? query returns the currently specified waveform
amplitude.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also
1094
(see
page
1208)
"Introduction to :WGEN<w> Commands"
":WGEN<w>:FUNCtion"
":WGEN<w>:VOLTage:OFFSet"
":WGEN<w>:VOLTage:HIGH"
":WGEN<w>:VOLTage:LOW"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
on page 1058
on page 1067
on page 1097
on page 1095
on page 1096