Wgen<W>:Track - Agilent Technologies InfiniiVision 4000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

34
:WGEN<w> Commands
:WGEN<w>:TRACk
Command Syntax
:WGEN<w>:TRACk <on_off>
<w> ::= 1 or 2 in NR1 format
<on_off> ::= {{1 | ON} | {0 | OFF}
The :WGEN<w>:TRACk ON command causes frequency, amplitude, offset,
and duty cycle adjustments to this waveform generator output signal to be
tracked by the other waveform generator output.
Not all waveform shapes can be frequency tracked. When
:WGEN<w>:TRACk or :WGEN<w>:TRACk:FREQuency is enabled, the other
generator's waveform selections (:WGEN<w>:FUNCtion) must be in the
same group as this generator's waveform:
Group 1
SINusoid
RAMP
SINC
CARDiac
GAUSsian
When tracking is enabled, adjustments for tracked settings in the other
waveform generator become unavailable.
Query Syntax
:WGEN<w>:TRACk?
The :WGEN<w>:TRACk? query returns whether tracking is on or off.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
1090
(see
page
1208)
Group 2
SQUare
PULSe
":WGEN<w>:FUNCtion"
":WGEN<w>:TRACk:AMPLitude"
":WGEN<w>:TRACk:FREQuency"
":WGEN<w>:TRACk:CSIGnal"
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
Group 3
EXPRise
EXPFall
on page 1067
on page 1091
on page 1093
on page 1092
Group 4
ARBitrary

Advertisement