:TIMebase:REFerence

Command Syntax
:TIMebase:REFerence <reference>
<reference> ::= {LEFT | CENTer | RIGHt}
The :TIMebase:REFerence command sets the time reference to one division
from the left side of the screen, to the center of the screen, or to one
division from the right side of the screen. Time reference is the point on
the display where the trigger point is referenced.
Query Syntax
:TIMebase:REFerence?
The :TIMebase:REFerence? query returns the current display reference for
the main window.
Return Format
<reference><NL>
<reference> ::= {LEFT | CENT | RIGH}
See Also
Example Code
t.
center.
See complete example programs at:
starting on page 1217
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
(see
page
1208)
"Introduction to :TIMebase Commands"
":TIMebase:MODE"
on page 925
' TIME_REFERENCE - Possible values are LEFT, CENTer, or RIGHt.
'
- LEFT sets the display reference one time division from the left.
'
- CENTer sets the display reference to the center of the screen.
'
- RIGHt sets the display reference one time division from the righ
myScope.WriteString ":TIMebase:REFerence CENTer"
:TIMebase Commands
on page 924
' Set reference to